Dimension Thyroid Stimulating Hormone (TSHL) Flex reagent cartridge
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Thyroid Stimulating Hormone (TSHL) Flex reagent cartridge on . FDA classified it as Class II and gives the reason as "The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1668-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Thyroid Stimulating Hormone (TSHL) Flex reagent cartridge
- Product code
- JLW Radioimmunoassay, Thyroid-Stimulating Hormone
- Reason for recall
- The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA
- Distribution
- Nationwide and Canada, Mexico, and Thailand
- Product codes and lots
- TSHL, radioimmunoassay, thyroid-stimulating hormone, RF612, SMN# 10464524 Lot codes: EB7276 EC7298 EA7318 EB7346 EC7346 EA8012 EA8033 EA8051 EC8080 EB8107 EB8121 EA8143 EB8143 EA8177 EB8177 EB8199 EA8223 EA8241 EA8261
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1668-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1669-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1670-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1671-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1672-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1673-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1674-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1675-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1676-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1677-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1678-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1679-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1680-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1681-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1682-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1683-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1684-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1685-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1686-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1687-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1688-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1689-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1690-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1691-2018 | Class II | Terminated | Voluntary: Firm initiated |