Dimension Vista Thyroid Stimulating Hormone (TSH) Flex reagent cartridge
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista Thyroid Stimulating Hormone (TSH) Flex reagent cartridge on . FDA classified it as Class II and gives the reason as "The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1673-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista Thyroid Stimulating Hormone (TSH) Flex reagent cartridge
- Product code
- JLW Radioimmunoassay, Thyroid-Stimulating Hormone
- Reason for recall
- The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA
- Distribution
- Nationwide and Canada, Mexico, and Thailand
- Product codes and lots
- TSH, radioimmunoassay, thyroid stimulating hormone, RF612, SMN# 10464524 Lot codes: 16277BD 16322BB 17005BB 17023AC 17069AA 17093BB 17122AA 17150AA 17193BA 17215AC 17268AA
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1668-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1669-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1670-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1671-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1672-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1673-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1674-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1675-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1676-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1677-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1678-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1679-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1680-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1681-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1682-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1683-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1684-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1685-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1686-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1687-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1688-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1689-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1690-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1691-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
