Z-1261-2018Class II

Syngo.via software is intended to be used for viewing, manipulation, communication, and storage of medical...

Siemens Medical Solutions USA, Inc. / initiated 2018-01-10 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Syngo.via software is intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a standalone device or... on . FDA classified it as Class II and gives the reason as "When the archiving configuration is changed, data received/created after the upgrade may be flagged as "Not to be archived". This is caused by the automatic function for cleaning up temporary data being disabled by the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1261-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo.via software is intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a standalone device or together with a variety of cleared and unmodified syngo based software options.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
When the archiving configuration is changed, data received/created after the upgrade may be flagged as "Not to be archived". This is caused by the automatic function for cleaning up temporary data being disabled by the software upgrade. Due to the disabled cleanup function, disc capacity for free space decreases faster than usual. Unless the archiving configuration and the cleanup automatic function are reverted back to the original state (prior to the software upgrade), data that was incorrectly flagged "Not to be archived" must be manually prevented from being deleted by either clinical administrator or service engineer.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
Customers received letters with information related to the issue and regarding the upgrade to the software version VB20A_HF04. The software was released December 22, 2017 and is available for install.
Quantity in commerce
165 units
Distribution
Medical device software which needs to be installed.
Product codes and lots
All Syngo.via systems with VB20A. Model number 10496180.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1261-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1261-2018Class IITerminatedVoluntary: Firm initiated