510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1317-2018Class II

D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius...

Smith & Nephew, Inc. / initiated 2018-03-07 / Terminated

Smith & Nephew, Inc. began a recall of D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF... on . FDA classified it as Class II and gives the reason as "One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1317-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R
Product code
HWC Screw, Fixation, Bone
Reason for recall
One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.
FDA root cause
Under Investigation by firm
Action
The firm initiated their recall by letter and email on 03/07/2018. The firm requested return of the product.
Quantity in commerce
59 units
Distribution
US, Puerto Rico, Malaysia
Product codes and lots
Lot Number 17GM04522

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1317-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1317-2018Class IITerminatedVoluntary: Firm initiated