D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius...
Smith & Nephew, Inc. began a recall of D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF... on . FDA classified it as Class II and gives the reason as "One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1317-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.
- FDA root cause
- Under Investigation by firm
- Action
- The firm initiated their recall by letter and email on 03/07/2018. The firm requested return of the product.
- Quantity in commerce
- 59 units
- Distribution
- US, Puerto Rico, Malaysia
- Product codes and lots
- Lot Number 17GM04522
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1317-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY