Z-1513-2018Class II

FilmArray BCID Panel (Blood Culture Identification Panel), Catalog numbers RFIT-ASY-0126 (30 test kit) and...

BioFire Diagnostics, LLC / initiated 2018-03-14 / Terminated / root cause: manufacturing process

BioFire Diagnostics, LLC began a recall of FilmArray BCID Panel (Blood Culture Identification Panel), Catalog numbers RFIT-ASY-0126 (30 test kit) and RFIT-ASY-0127 (6 test kit), when used with all lots... on . FDA classified it as Class II and gives the reason as "There is an increased risk of false positive Proteus results when the product is used with specific types of BD BACTEC blood culture bottles". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1513-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BioFire Diagnostics, LLC
Product
FilmArray BCID Panel (Blood Culture Identification Panel), Catalog numbers RFIT-ASY-0126 (30 test kit) and RFIT-ASY-0127 (6 test kit), when used with all lots of BD BACTEC Blood Culture Bottles, Catalog numbers 442020, 442021, 442023, 442192, 442265, 442194, 442022, 442193, 442260, and 442191.
Product code
PEN Gram-Negative Bacteria And Associated Resistance Markers
Reason for recall
There is an increased risk of false positive Proteus results when the product is used with specific types of BD BACTEC blood culture bottles.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
The recalling firm initiated the recall on several different dates and methods: Via email on 3/14/2018 with the letter dated 3/3/2018 attached, via USPS on 3/23/2018, and via FedEx between 3/24/2018 and 3/28/2018. The recalling firm conducted an expansion and issued letters dated 6/11/2018 via FedEx or USPS on 6/13/2018.
Distribution
Distribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Algeria, Argentina, Australia, Austria, Bahrain, Botswana, Brazil, Chile, Colombia, Costa Rica, Czech Republic, Dominican Republic, Ecuador, Egypt, France, French Polynesia, Georgia, Germany, Greece, Guatemala, Hong Kong, India, Iraq, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Lebanon, Malaysia, Mexico, Myanmar, The Netherlands, New Caledonia, Nicaragua, Panama, Peru, Republic of the Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, Uganda, United Arab Emirates, United Kingdom, Uruguay.
Product codes and lots
All lot numbers are involved. UDI: 00815381020086 (RFIT-ASY-0126), 00815381020093 (RFIT-ASY-0127).  The expiration dates of the BD BACTEC Blood Culture Bottles are 9/30/2018 and beyond.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1513-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1513-2018Class IITerminatedVoluntary: Firm initiated