ARROWg+ard Blue PLUS catheter, Catalog Number: ASK-45854-NG
Arrow International Inc began a recall of ARROWg+ard Blue PLUS catheter, Catalog Number: ASK-45854-NG Product Usage: The ARROWg+ard Blue PLUS catheter permits venous access to central circulation by... on . FDA classified it as Class II and gives the reason as "Certain lots may be missing information at the bottom of the lid including the lot number, expiration date, and Unique Device Identification (UDI) information". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1152-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- ARROWg+ard Blue PLUS catheter, Catalog Number: ASK-45854-NG Product Usage: The ARROWg+ard Blue PLUS catheter permits venous access to central circulation by way of subclavian, jugular, and femoral veins. The ARROWg+ard technology is intended to provide protection against catheter-related bloodstream infections (CRBSIs). It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use (> 30 days). Clinical effectiveness of the ARROWg+ard Blue PLUS catheter in preventing CRBSIs compared to the original ARROWg+ard Blue catheter has not been studied.
- Product code
- FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
- Reason for recall
- Certain lots may be missing information at the bottom of the lid including the lot number, expiration date, and Unique Device Identification (UDI) information.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Arrow International sent Notification letters to DISTRIBUTORS that include the following instructions: 1. Discontinue distribution and quarantine any affected products. 2. Inspect affected products to identify whether they have the lot number, expiration date, and UDI information. Products missing this information should be returned. 3. Using the provided Customer Letter and Acknowledgement Form templates, communicate this recall to any customers who have received affected products. Customers should return mislabeled products. Products that are properly labeled should not be returned. 4. Any customers with mislabeled products should return them with a completed Acknowledgement Form. 6. If you and customers have no affected stock please complete and return the Acknowledgment Form. Notification letters to RISK MANAGERS include the following instructions: 1. Discontinue use and quarantine any affected products. 2. Inspect affected products and identify whether they have the lot number, expiration date, and UDI information. Products missing this information should be returned to the Distributor. 3. Complete the Acknowledgement Form whether or not there is affected product that must be returned. For questions contact your local sales representative or Customer Service at 1-866-246-6990.
- Quantity in commerce
- 420
- Distribution
- United States in the state of FL.
- Product codes and lots
- Lot Numbers: 23F16H0744, 23F16J0072, 23F17B0781