DS Titanium Ligation Clip, used in laparoscopic surgery
Aesculap Implant Systems LLC began a recall of DS Titanium Ligation Clip, used in laparoscopic surgery on . FDA classified it as Class II and gives the reason as "Added a contraindication to IFU: Do not use the DS clips for living donor nephrectomies. Added other contraindications. A publication from 2017 recommended Aesculap DS Clips for kidney transplantation with living...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1290-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Aesculap Inc
- Product
- DS Titanium Ligation Clip, used in laparoscopic surgery
- Product code
- FZP Clip, Implantable
- Reason for recall
- Added a contraindication to IFU: Do not use the DS clips for living donor nephrectomies. Added other contraindications. A publication from 2017 recommended Aesculap DS Clips for kidney transplantation with living donors. The information from this publication was not approved or validated by the firm.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- A letter and copy of the updated IFU was sent to affected customers the week of January 15, 2018. Product does not need to be returned. IFU is revised and put on the Aesculap Website for usage. https://www.aesculapusa.com/products/instructions-for-use.
- Quantity in commerce
- 9621
- Distribution
- US and Canada distribution
- Product codes and lots
- DS Clips - Product Numbers PL450SU DS SINGLE FIRE LIG.CLIPS PL471SU DS SINGLE FIRE LIG.CLIP L W/LATCH PL453SU DS SINGLE FIRE LIG.CLIP SM PL475SU DS SINGLE FIRE LIG.CLIP XL W/LATCH PL459SU DS SINGLE FIRE LIG.CLIP M PL462SU DS SINGLE FIRE LIG.CLIP ML PL465SU DS SINGLE FIRE LIG.CLIP ML W/LATCH PL468SU DS SINGLE FIRE LIG.CLIP L (all clips going back 5 years)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1290-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1291-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY