Z-1148-2018Class II

Artis one fluoroscopic x-ray system and patient table Artis one is an angiography system developed for...

Siemens Medical Solutions USA, Inc. / initiated 2018-01-12 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Artis one fluoroscopic x-ray system and patient table Artis one is an angiography system developed for diagnostic imaging and interventional procedures... on . FDA classified it as Class II and gives the reason as "Certain patient table systems may have a missing or incorrectly installed snap ring that could cause the bolt to slide out of its base and the tabletop carriage to slide to the floor". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1148-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis one fluoroscopic x-ray system and patient table Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Arts one include cardiac angiography, neuro-angiography, general angiography, rotational angiography; multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Certain patient table systems may have a missing or incorrectly installed snap ring that could cause the bolt to slide out of its base and the tabletop carriage to slide to the floor.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Siemens sent an Urgent - Medical Device Correction letter dated January 5, 2018. Siemens will conduct an inspection of all affected Artis tables and verify the correct positioning of the bolt as well as the snap rings. The letter recommends the consignees follow Please follow the standard emergency procedures they have in place in the event of system failures. It is recommended to consider reinforcing these emergency procedures until the update has been completed. The Siemens service organization will contact the consignees to arrange a date to perform the corrective action. For further questions, please call (610) 448-6461.
Quantity in commerce
16
Distribution
US Distribution to the states of : AL, AR, CA, FL, IN, LA, MI, MO, NC, NY, PA, TX, and WA.
Product codes and lots
Serial numbers: 82044 82105 82070 82213 82054 82040 82038 82205 82065 82267 82108 82010 82039 82109 82102 82066

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1148-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1148-2018Class IITerminatedVoluntary: Firm initiated