Z-1138-2018Class II
MyLab Ultrasound System; Models 6100, 6150, 6200, 6250; Probes: BL433, C 18, CA541, EC1123, EC123, IH 618...
Esaote S.p.A. / initiated 2018-01-03 / Terminated / root cause: software
ESAOTE S.P.A. began a recall of MyLab Ultrasound System; Models 6100, 6150, 6200, 6250; Probes: BL433, C 18, CA541, EC1123, EC123, IH 618, IOT332, IOT342, L 415, LA332E, LA435, LA523, LA533... on . FDA classified it as Class II and gives the reason as "The ultrasound system used in certain settings may result in overheating of the ultrasound probe head". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1138-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Esaote S.p.A.
- Product
- MyLab Ultrasound System; Models 6100, 6150, 6200, 6250; Probes: BL433, C 18, CA541, EC1123, EC123, IH 618, IOT332, IOT342, L 415, LA332E, LA435, LA523, LA533, SI2C41, SL2325, TRT33
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- The ultrasound system used in certain settings may result in overheating of the ultrasound probe head.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The January 3, 2018, notification letter requests the consignee complete and return a recall response form. The letter states that Esaote customer service will contact the consignee about the software update. In the meantime, it is recommended that the consignee does not use the PA023E probe in Pulsed Wave (PW) Doppler mode and that they refrain from using the system in Elaxto mode where the patient is unconscious or sedated. The consignee is asked to check the temperature of the probe before use.
- Quantity in commerce
- 5
- Distribution
- US distribution in the states: AZ, FL, OK, PA, and TX.
- Product codes and lots
- Software versions 12.10 - 13.60
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1137-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1138-2018 | Class II | Terminated | Voluntary: Firm initiated |