Z-1320-2018Class II

Hematology Diluent M-30D used with BC-3200 and BC-3600 Hematology Analyzer; Part Numbers: (1) 105-005405-00...

Datascope Corp. / initiated 2018-01-24 / Terminated / root cause: manufacturing process

Mindray DS USA, Inc. dba Mindray North America began a recall of Hematology Diluent M-30D used with BC-3200 and BC-3600 Hematology Analyzer; Part Numbers: (1) 105-005405-00 (2 x 5.5L container), (2) 105-005406-00 (20L... on . FDA classified it as Class II and gives the reason as "Certain lots of M-30D diluent used with BC-3200 and BC-3600 Hematology Analyzers and M-53D diluent used with BC-5390 Hematology Analyzers may cause the system to produce an elevated platelet background count". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1320-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Hematology Diluent M-30D used with BC-3200 and BC-3600 Hematology Analyzer; Part Numbers: (1) 105-005405-00 (2 x 5.5L container), (2) 105-005406-00 (20L container) The M-30D Diluent is an azide-free, filtered isotonic solution for counting and sizing blood cells. It is for use with the Mindray BC-3200 and BC-3600 Hematology Analyzer.
Product code
GIF Diluent, Blood Cell
Reason for recall
Certain lots of M-30D diluent used with BC-3200 and BC-3600 Hematology Analyzers and M-53D diluent used with BC-5390 Hematology Analyzers may cause the system to produce an elevated platelet background count.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Letters dated January 24, 2018, were sent to consignees with the following instructions: The issue manifests intermittently, therefore, you may continue to use affected lots if the background check is successfully completed. Discard affected lots which cause the system to display the error message background abnormal. Mindray will replace the potentially affected lots at no cost. Please contact Technical Support to order replacement product. For further questions, please call (201) 995-8407.
Quantity in commerce
4838
Distribution
US Distribution including Puerto Rico and to the states of : CA, MI, NC, TX, and WA. .
Product codes and lots
(1) 105-005405-00 (2 x 5.5L container) Lot Numbers: 2017060906, 2017041207, 2017042707, 2017041104, 2016122301, 2016112901, 2016060401, 2016052701, 2016052101, 2016050301; (2)105-005406-00 (20L container) Lot Numbers: 2017041104, 2016121901, 2016051801, 2016050301

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1320-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1320-2018Class IITerminatedVoluntary: Firm initiated
Z-1321-2018Class IITerminatedVoluntary: Firm initiated