Z-1200-2018Class II
ACUSON SC2000 Ultrasound System. The firm name on the label is Siemens Medical Solutions USA, Inc., Buffalo...
Siemens Medical Solutions USA, Inc. / initiated 2018-03-05 / Terminated / root cause: design
Siemens Medical Solutions USA, Inc. began a recall of ACUSON SC2000 Ultrasound System. The firm name on the label is Siemens Medical Solutions USA, Inc., Buffalo Grove, IL on . FDA classified it as Class II and gives the reason as "The ECG signal may flatline due to electromagnetic interference during the use of electrosurgical equipment". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1200-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ACUSON SC2000 Ultrasound System. The firm name on the label is Siemens Medical Solutions USA, Inc., Buffalo Grove, IL.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- The ECG signal may flatline due to electromagnetic interference during the use of electrosurgical equipment.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Undated recall letters were issued via email (if they had an email address for the customer) or via certified mail (if no email address was available) on 3/5/2018.
- Quantity in commerce
- 229 systems
- Distribution
- Distribution was made to medical facilities in AZ, CA, CT, FL, IL, IN, KS, KY, MI, MN, MO, NJ, NY, NC, OH, PA, TX, VA, and WA. There was no military/government distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Bosnia Herzeg, Colombia, Croatia, Denmark, France, Germany, India, Ireland, Israel, Italy, Japan, Mexico, New Zealand, Philippines, Poland, Republic Korea, Romania, Singapore, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, and Vietnam.
- Product codes and lots
- Software versions VB20A, VB20B, or VB20C that have a common physio module (CPM) (known by customers as the ECG (physio) hardware) installed at revision 4
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1200-2018 | Class II | Terminated | Voluntary: Firm initiated |