Z-1275-2018Class II

Ray Station 4.9 Ray Station 5, Ray Station 6 and Ray Station 7 Software build numbers: 4.9.0.42, 5.0.0.37...

RaySearch Laboratories AB (publ) / initiated 2018-01-26 / Terminated / root cause: design

RAYSEARCH LABORATORIES AB began a recall of Ray Station 4.9 Ray Station 5, Ray Station 6 and Ray Station 7 Software build numbers: 4.9.0.42, 5.0.0.37, 5.0.1.11, 5.0.2.35, 6.0.0.24, 6.1.0.26, 6.1.1.2... on . FDA classified it as Class II and gives the reason as "Software issue with Center Beam in Field functionality. Issue can result in incorrect treatment volume delivered to patient". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1275-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
RaySearch Laboratories AB (publ)
Product
Ray Station 4.9 Ray Station 5, Ray Station 6 and Ray Station 7 Software build numbers: 4.9.0.42, 5.0.0.37, 5.0.1.11, 5.0.2.35, 6.0.0.24, 6.1.0.26, 6.1.1.2, 6.2.0.7 or 7.0.0.19 UDI: 0735000201006820171130
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Software issue with Center Beam in Field functionality. Issue can result in incorrect treatment volume delivered to patient.
Root cause tag
design
FDA root cause
Device Design
Action
On January 26, 2018 RaySearch Laboratories AB sent Field Safety Notice, Medical Device Correction #25955 to each customer via e-mail. Customers were advised of the following: be aware that Center Beam in Field affects all beams that share the same isocenter and that apertures may be modified in a way not intended; do not use Center Beam in Field after having created an opposed beam. If an opposed beam has been created, delete the opposed beam and use Center Beam in Field on the remaining beam. Then recreate the opposed beam; please educate planning staff and all users about this workaround; inspect your product and identify all installed units with the above software version number(s), then confirm you have read and understood this notice by replying to the notification via email, fax or by calling local support. Customers with questions were instructed to contact David Hedfors via e-mail at [email protected]
Quantity in commerce
2484
Distribution
Nationwide
Product codes and lots
Software build numbers: 4.9.0.42, 5.0.0.37, 5.0.1.11, 5.0.2.35, 6.0.0.24, 6.1.0.26, 6.1.1.2, 6.2.0.7 or 7.0.0.19

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1275-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1275-2018Class IITerminatedVoluntary: Firm initiated