Z-1644-2018Class II
RESTORIS MCK TIBIAL BASEPLATE-RM/LL used in unicompartmental knee arthroplasty and knee replacements
Mako Surgical Corp. / initiated 2017-12-01 / Terminated / root cause: labeling and UDI
Mako Surgical Corporation began a recall of RESTORIS MCK TIBIAL BASEPLATE-RM/LL used in unicompartmental knee arthroplasty and knee replacements on . FDA classified it as Class II and gives the reason as "Incorrect product and/or label. Sizing is labeled incorrectly". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1644-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mako Surgical Corp.
- Product
- RESTORIS MCK TIBIAL BASEPLATE-RM/LL used in unicompartmental knee arthroplasty and knee replacements.
- Product code
- OLO Orthopedic Stereotaxic Instrument
- Reason for recall
- Incorrect product and/or label. Sizing is labeled incorrectly.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- An URGENT MEDICAL DEVICE RECALL was sent on December 05, 2017 Product Field Action #: 1658081 Description: MCK TIBIAL BASEPLATE-RM/LL-SZ 2 and MCK TIBIAL BASEPLATE-RM/LL-SZ 7 Item No.: 180612; Lot Number: 26080317-01180617; and Lot Number: 26150217-01 Dear Customer, Stryker has initiated an urgent, voluntary, lot-specific recall for the MCK TIBIAL BASEPLATE-RM/LL-SZ 2 and MCK TIBIAL BASEPLA TE-RM/LL-SZ 7 referenced above. The intent of this letter is to inform you of the product recall that was initiated on December 01, 2017 by Stryker, and to list known hazards potentially associated with the use of the above referenced products and list the risk mitigation factors. Stryker has discovered that the packaging of certain sizes and lots of the above-referenced product may contain the incorrect product and/or label. Two reports were received with the product/label discrepancy. In one report, the labeling of the implant box outer label stated Size 2 RM/LL, and the labeling of the implant sticker (Patient label) located inside the outer box stated Size 7 RM/LL. The correct implant Size 2RM/LL was inside the box. The patient label was incorrect in this report. The second report described that a size 2 implant was in a box labeled as a size 7 implant. Potential Hazards Technical and medical assessments are currently underway to determine any potential hazards associated with the use of the product. Additional communication will be forwarded upon completion of the internal investigation on this issue. Risk Mitigation The difference in Size 2 RM/LL implant and Size 7 RM/LL implant is easily identified by the end user and would not likely be implanted as the discrepancy would be obvious. Actions Needed 1. Please inform users of this Urgent Product Recall and forward this notice to all individuals who need to be aware within your organization 2. Hospitals/Branches/Agencies: Complete and sign the enclosed Recall Notification Business Reply Form and
- Quantity in commerce
- 126
- Distribution
- AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, LA, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OK, OR, PA, RI, SC, TX, UT, VA, WA, WY, and Hawaii Australia, Italy, Netherlands, Singapore, South Korea, United Kingdom, and Vietnam
- Product codes and lots
- Size 2 Item # 180612 Lot # 26080317-01 and Size 7 Item # 180617 Lot # 26150217-01.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1644-2018 | Class II | Terminated | Voluntary: Firm initiated |