Z-1478-2018Class II

Flowmeter Module (accessory to Terumo Advanced Perfusion System 1). Provides the interface between the flow...

Terumo Cardiovascular Systems Corporation / initiated 2018-01-19 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of Flowmeter Module (accessory to Terumo Advanced Perfusion System 1). Provides the interface between the flow sensor and the system on . FDA classified it as Class II and gives the reason as "Inaccurrate flow readings. Depending on the degree of inaccuracy, this issue may not be easy for the user to detect during setup or use (for example, following a Flow Probe relocation or manipulation)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1478-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Flowmeter Module (accessory to Terumo Advanced Perfusion System 1). Provides the interface between the flow sensor and the system.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Inaccurrate flow readings. Depending on the degree of inaccuracy, this issue may not be easy for the user to detect during setup or use (for example, following a Flow Probe relocation or manipulation).
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The firm notified customers via phone alerting them of this affected device, issue, potential hazard, correction, and instructions. The communication also included scheduling an expedited service call for a Field Service Technician to replace the affected Flowmeter Module with a corrected Flowmeter Module. When necessary to avoid delaying or cancelling life-sustaining surgery, users can continue to use the Flowmeter Module while awaiting replacement. Once the issue is recognized by the user, if a replacement Flowmeter Module is available, replacement and reassignment of safety connections of the Flowmeter Module can be accomplished in less than 15 seconds. In the event that a replacement Flowmeter Module is not available, a less common mitigation is the use of a back-up stand-alone centrifugal pump or a stand-alone ultrasonic flowmeter system to provide flow data. Customers should receive a copy of the Urgent Medical Device Recall phone script by e-mail along with a Customer Response Form. The e-mail should be reviewed, and the form completed and returned as indicated. Questions or concerns can be directed Terumo CVS Customer Service at 1-800-521-2818.
Quantity in commerce
7
Distribution
Worldwide Distribution -- USA, including the states of IL, KY, and MI; and, the country of Singapore.
Product codes and lots
Catalog # - 802018 Serial #: 01662, 01672, 01680, 01681,01682, 01683 , and 01687. UDI: 00886799000687.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1478-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1478-2018Class IITerminatedVoluntary: Firm initiated