Z-1798-2018Class II
FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2), Laboratory Automation System
Inpeco S.p.A. / initiated 2018-02-13 / Terminated / root cause: design
Inpeco S.A. began a recall of FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2), Laboratory Automation System on . FDA classified it as Class II and gives the reason as "Module may freeze without generating user warning. There is a potential risk in delay of sample processing, leading to delayed delivery of test results to patients". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1798-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Inpeco S.p.A.
- Product
- FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2), Laboratory Automation System
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- Module may freeze without generating user warning. There is a potential risk in delay of sample processing, leading to delayed delivery of test results to patients.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 2/13/18, distributors were notified by emailed Technical Service Bulletin. The recalling firm also released a Customer Letter to be forwarded to final customers (labs) by distributors. The Customer Letter advised users of the issue and informed that the recalling firm would be visiting sites to upgrade the firmware to resolve the issue. All impacted customers were instructed to complete and return the Receipt Confirmation within 7 days. Contact provided: Giorgia Amabile - Quality Assurance & Regulatory Affairs Manager, [email protected], (+41) 91 9118 258.
- Quantity in commerce
- 18
- Distribution
- Distributors in 2 states: NY and IL.
- Product codes and lots
- Serial Numbers: Accelerator a3600 - ACP.150 ACP.185 ACP.186 ACP.192 ACP.202 ACP.214 FlexLab - FLX.0165 FLX.0127 FLX.0128 FLX.0133 FLX.0157 FLX.0160 FLX.0181 Aptio Automation - AP2.0054 AP2.0056 AP2.0060 AP2.0078 AP2.0095 Part Numbers of impacted sample tube carriers: 721501001 for 43 mm carriers, and 0A00006957 for 40 mm carriers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1798-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
