Intra-Aortic Balloon Catheter Mega 8FR 50cc
Datascope Corporation began a recall of Intra-Aortic Balloon Catheter Mega 8FR 50cc Product Usage: The intra-aortic balloon catheter and accessories are used to provide counter pulsation therapy in... on . FDA classified it as Class II and gives the reason as "The U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1385-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- Intra-Aortic Balloon Catheter Mega 8FR 50cc Product Usage: The intra-aortic balloon catheter and accessories are used to provide counter pulsation therapy in the aorta, whereby balloon inflation during diastole and deflation during systole increases blood supply to the heart muscle and decreases the work of the left ventricle.
- Product code
- DSP System, Balloon, Intra-Aortic And Control
- Reason for recall
- The U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Maquet/Getinge issued a recall letter on February 21, 2018 to affected customers, via FEDEX with confirmation of delivery, to notify them of the fact that the IFU is missing the Contraindications section and to provide the missing information. Affected customers are instructed to examine their inventory immediately to determine if they have affected product. An Addendum to the Mega IAB Catheter IFU, containing the Contraindications, is provided as an attachment in the recall letter. It is suggested that affected customers post copies of the Addendum near all affected lot numbers of Mega IAB Catheters. Customers are provided with Maquet/Getinge contact information should their facility require a full copy of the correct Mega IAB Catheter IFU. Lastly, affected customers are instructed to complete and sign a response form, acknowledging that they have received the notification, and return the form to Maquet/Getinge.
- Quantity in commerce
- 12744
- Distribution
- US Nationwide Distribution
- Product codes and lots
- 0684 00 0296 01U and 0684 00 0296 02U
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1383-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1384-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1385-2018 | Class II | Terminated | Voluntary: Firm initiated |