Edwards Lifesciences' IntraClude Intra-Aortic Occlusion Device. Packaged in a sterile barrier system. Model...
Edwards Lifesciences, LLC began a recall of Edwards Lifesciences' IntraClude Intra-Aortic Occlusion Device. Packaged in a sterile barrier system. Model Number: ICF100. Designed to occlude the ascending... on . FDA classified it as Class II and gives the reason as "Possible leak at distal open end of the strain relief, which causes medical staff to continue adding inflation fluid to maintain balloon pressure for adequate occlusion of the ascending aorta". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1929-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Edwards Lifesciences LLC
- Product
- Edwards Lifesciences' IntraClude Intra-Aortic Occlusion Device. Packaged in a sterile barrier system. Model Number: ICF100. Designed to occlude the ascending aorta.
- Product code
- DXC Clamp, Vascular
- Reason for recall
- Possible leak at distal open end of the strain relief, which causes medical staff to continue adding inflation fluid to maintain balloon pressure for adequate occlusion of the ascending aorta.
- FDA root cause
- Under Investigation by firm
- Action
- Customers were notified on about 03/30/2018 via an Urgent Medical Device Safety Notice. Instructions included to review the field safety notice and understand the potential hazard, review the recall with appropriate clinical staff at their institution, complete and return the acknowledgement form, and distribute the notification within the organization where potentially affected devices may have been transferred. Questions can be directed to Edwards Customer Service, Monday through Friday at 800-424-3278 form 8:00 AM - 4:00 PM Pacific Time.
- Quantity in commerce
- 3,052 units total
- Distribution
- Worldwide Distribution -- USA, to the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, KY, MA, MI, MS, MO, NV, NJ, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, and WV; and, the countries of to Australia, Austria, Belgium, Canada, Colombia, Czech Republic, France, Germany, Israel, Italy, Luxembourg, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
- Product codes and lots
- Lot No. 60876468, 60850885, 6060850886, 60850888, 60850889, 60850890, 60887217, 60887218, 60887219, 60887220, 60930494, 60934623, 60934624, 60934625, 60972889, 60972890, 60972892, 61014406, 61014408, 61014410, 61014411, 61050171, 61050172, 61078031, 61078032, 61097633, 61139239, 61139240, 61139241, 61139242, 61184964, 61184965, 67784966, 61184967, 61204586, 61204587, 61259626, 61259627, 61259628, and 61259629.