510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1416-2018Class II

Terumo HX2 Temperature Management System, Catalog No. 809810 The Terumo HX2 Temperature Management System is...

Terumo Cardiovascular Systems Corporation / initiated 2018-03-16 / Terminated / root cause: labeling and UDI

Terumo Cardiovascular Systems Corporation began a recall of Terumo HX2 Temperature Management System, Catalog No. 809810 The Terumo HX2 Temperature Management System is indicated for use to supply temperature regulated... on . FDA classified it as Class II and gives the reason as "Update to cooler-heater cleaning instructions". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1416-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo HX2 Temperature Management System, Catalog No. 809810 The Terumo HX2 Temperature Management System is indicated for use to supply temperature regulated water to heat exchangers during open heart surgery.
Product code
DWC Controller, Temperature, Cardiopulmonary Bypass
Reason for recall
Update to cooler-heater cleaning instructions.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
The firm,Terumo Cardiovascular Systems Corporation, mailed customers an "Urgent Medical Device Correction" letter dated 2/16/18. The letter described the product, problem and actions to be taken. The Customers were instructed to do the following: 1) discard the current Sarns TCM, TCM II Cooling and Heating, or tHX2 Temperature Management Systems cleaning guides; 2) Review the notice and updated cleaning guides, and ensure all users receive notice of the updated cleaning guides and; 3) Return the Customer Response Form acknowledging receipt of the notice via Terumo Recall Email: [email protected] or Terumo Recall Fax: 734-741-6149. The letter also includes a list of the updates to the cleaning guides. Customers with questions may contact Terumo CVS Customer Service (800) 521-2818, Monday-Friday, 8am-6pm ET or Fax (734)741-6149.
Quantity in commerce
2393 this total is for all products
Distribution
Worldwide Distribution: US (Nationwide), Australia, Belgium, Canada, Chile, China. Colombia, Dominican Republic, England, Finland, France, Germany, Greece, HK, India, Indonesia, Iran. Israel, Italy, Japan, Korea, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates,and Vietnam.
Product codes and lots
All Lot Codes

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1416-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1416-2018Class IITerminatedVoluntary: Firm initiated
Z-1417-2018Class IITerminatedVoluntary: Firm initiated
Z-1418-2018Class IITerminatedVoluntary: Firm initiated