Z-1610-2018Class II

cobas b 221<2>Roche OMNI S2 system catalog numbers: 3337111001 and 3337111692 The Roche Diagnostics Omni S...

Roche Diagnostics GmbH / initiated 2018-02-22 / Terminated / root cause: software

Roche Diagnostics Corporation began a recall of cobas b 221<2>Roche OMNI S2 system catalog numbers: 3337111001 and 3337111692 The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer... on . FDA classified it as Class II and gives the reason as "The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1610-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
cobas b 221<2>Roche OMNI S2 system catalog numbers: 3337111001 and 3337111692 The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
Product code
CEM Electrode, Ion Specific, Potassium
Reason for recall
The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Root cause tag
software
FDA root cause
Software design
Action
ROCHE sent an Urgent Medical Device Correction Letter dated February 22, 2018, via UPS to affected customers. Customers were instructed to complete the following: " 1) Use the Return to Setup and Exit buttons to ensure execution of scheduled AutoQC measurements as described in this Urgent Medical Device Correction (UMDC); 2) Check to see if the next scheduled AutoQC measurement is performed to ensure the AutoQC scheduler is working as expected; 3) If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site; 4) Complete the attached fax form (TP-00146) and fax it to 1-844-491-7776 or scan and email to [email protected]; 5) File this UMDC for future reference. Customers with questions should contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272.
Quantity in commerce
1026 total products
Distribution
USA ( nationwide)
Product codes and lots
3337111001 and 3337111692  UDI: 04015630018307

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1610-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1610-2018Class IITerminatedVoluntary: Firm initiated
Z-1611-2018Class IITerminatedVoluntary: Firm initiated
Z-1612-2018Class IITerminatedVoluntary: Firm initiated
Z-1613-2018Class IITerminatedVoluntary: Firm initiated