IQon Spectral CT - Model no. 728332, Computed Tomography X-Ray system
Philips Medical Systems (Cleveland) Inc began a recall of IQon Spectral CT - Model no. 728332, Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional... on . FDA classified it as Class II and gives the reason as "Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1414-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- IQon Spectral CT - Model no. 728332, Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm notified customers of the recall on February 8, 2018, via Urgent Field Safety Notice. The notice advised customers of the issue and affected products and lot numbers. Customers were advised that once the E-stop has been closed, the operator/technologist must initate a brief horizontal motion (in or out) using any of the operator control panels, CT-Box or tape/foot switches. This action will ensure the couch carbon top (upper pallet) is in a controlled state. Customers were informed that a Field Service Engineer would contact them to schedule site updates to correct the issue. Customers who need further information or support were advised to contact their local Philips representative. Additionally, they may contact the Customer Care Solutions Center (1-800-722-9377.
- Quantity in commerce
- 47
- Distribution
- Worldwide Distribution - U.S. Nationwide. International distribution to Canada, Mexico, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Chile, China, Colombia, Croatia, Cuba, Czech Republic, Denmark, Djibouti, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Ghana, Gibraltar, Greece, Guyana, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Laos, Lebanon, Lithuania, Macedonia, Malaysia, Maldives, Mali, Martinique, Monaco, Mongolia, Myanmar, Nepal, Netherlands, New Calcedonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Rwanda, Saint Pierre and Miquelon, Samoa, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, and Vietnam.
- Product codes and lots
- Serial no: 860023 860035 860047 860059 860010 860014 860015 860013 860029 860006 860002 860061 860011 860020 860024 860008 860039 860043 860018 860012 860022 860032 860033 860040 860009 860016 860017 860042 860003 860005 860037 860041 860025 860038 860036 860052 860056 860021 860053 860019 860054 860044 860028 860030 860026 860031 860034
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1404-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1405-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1406-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1407-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1408-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1409-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1410-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1411-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1412-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1413-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1414-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1415-2018 | Class II | Terminated | Voluntary: Firm initiated |