510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1736-2018Class II

Gentle Threads, Interference Screw; Model No. 905605, for soft tissue reattachment procedures in the...

Zimmer GmbH / initiated 2018-03-28 / Terminated / root cause: manufacturing process

Zimmer Biomet, Inc. began a recall of Gentle Threads, Interference Screw; Model No. 905605, for soft tissue reattachment procedures in the ankle/foot and knee on . FDA classified it as Class II and gives the reason as "Specific lots of the device were overexposed during EtO sterilization. This may lead to revision due to infection from loss of sterile barrier during over-exposure cycle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1736-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Gentle Threads, Interference Screw; Model No. 905605, for soft tissue reattachment procedures in the ankle/foot and knee.
Product code
HWC Screw, Fixation, Bone
Reason for recall
Specific lots of the device were overexposed during EtO sterilization. This may lead to revision due to infection from loss of sterile barrier during over-exposure cycle.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The recalling firm issued letters to Distributors as well as to affected Risk Managers and Surgeons. Distributors were instructed to ensure affected team members are aware of the contents of the recall notification, locate and quarantine affected product in inventory, return all affected product from distributorship and from affected hospitals, and submit a list of any additional customer accounts that were not notified of the recall to the recalling firm. Risk Managers were instructed to ensure affected personnel are aware of the contents of the recall notification, quarantine all affected product, and complete the included Certificate of Acknowledgement and return to the recalling firm, even if no affected products were present at the facility. Surgeons were instructed to review the notification and return the completed Certificate of Acknowledgement. Customers' questions and concerns were directed to the recalling firm's customer service at 574-371-3071, 8AM-5PM EST, Monday through Friday. Alternatively, questions may be emailed to [email protected].
Quantity in commerce
53
Distribution
Worldwide distribution. US nationwide, Netherlands, Mexico, Paraguay, Brazil, Argentina, Malaysia, Costa Rica, Chile, South Korea, and Canada.
Product codes and lots
lot 326900

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1736-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1734-2018Class IITerminatedVoluntary: Firm initiated
Z-1735-2018Class IITerminatedVoluntary: Firm initiated
Z-1736-2018Class IITerminatedVoluntary: Firm initiated
Z-1737-2018Class IITerminatedVoluntary: Firm initiated
Z-1738-2018Class IITerminatedVoluntary: Firm initiated
Z-1739-2018Class IITerminatedVoluntary: Firm initiated