Z-1587-2018Class II

Atellica Solution and Atellica Sample Handler Prime for in vitro diagnostic testing of clinical specimens

Siemens Healthcare Diagnostics Inc. / initiated 2018-02-26 / Terminated / root cause: software

Siemens Healthcare Diagnostics, Inc. began a recall of Atellica Solution and Atellica Sample Handler Prime for in vitro diagnostic testing of clinical specimens on . FDA classified it as Class II and gives the reason as "When more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1587-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Atellica Solution and Atellica Sample Handler Prime for in vitro diagnostic testing of clinical specimens.
Product code
JGS Electrode, Ion Specific, Sodium
Reason for recall
When more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers
Root cause tag
software
FDA root cause
Software design
Action
Recall initiated 02/23/2018 Firm released an update software (SW) version (V), 1.12.1, which corrects the behavior. A Mandatory Update (MU) campaign was launched to install the new SW V on all systems, whether or not they are impacted by this behavior. The MU campaign will be tracked to 100% completion globally.
Distribution
FL and France Spain Sweden United Kingdom Germany
Product codes and lots
UDI 00630414002439

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1587-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1586-2018Class IITerminatedVoluntary: Firm initiated
Z-1587-2018Class IITerminatedVoluntary: Firm initiated
Z-1588-2018Class IITerminatedVoluntary: Firm initiated