Analyzer and Atellica IM 1600 Analyzer system for in vitro diagnostic testing of clinical specimens
Siemens Healthcare Diagnostics, Inc. began a recall of Analyzer and Atellica IM 1600 Analyzer system for in vitro diagnostic testing of clinical specimens on . FDA classified it as Class II and gives the reason as "When more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1588-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Analyzer and Atellica IM 1600 Analyzer system for in vitro diagnostic testing of clinical specimens.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- When more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Recall initiated 02/23/2018 Firm released an update software (SW) version (V), 1.12.1, which corrects the behavior. A Mandatory Update (MU) campaign was launched to install the new SW V on all systems, whether or not they are impacted by this behavior. The MU campaign will be tracked to 100% completion globally.
- Distribution
- FL and France Spain Sweden United Kingdom Germany
- Product codes and lots
- UDI 00630414002026
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1586-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1587-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1588-2018 | Class II | Terminated | Voluntary: Firm initiated |