510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1522-2018Class II

Orthopedic Salvage System (OSS) Modular Arthrodesis 0 Degree Locking Collar, Model Number CP260600

Zimmer GmbH / initiated 2018-04-03 / Terminated / root cause: manufacturing process

Zimmer Biomet, Inc. began a recall of Orthopedic Salvage System (OSS) Modular Arthrodesis 0 Degree Locking Collar, Model Number CP260600 Product Usage: The replacement of segmental defects of long... on . FDA classified it as Class II and gives the reason as "There is a potential dry blast irregularity on the locking collar screws packaged with the locking collar". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1522-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Orthopedic Salvage System (OSS) Modular Arthrodesis 0 Degree Locking Collar, Model Number CP260600 Product Usage: The replacement of segmental defects of long bone, including midshaft replacement, and arthrodesis of the knee.
Product code
HSB Rod, Fixation, Intramedullary And Accessories
Reason for recall
There is a potential dry blast irregularity on the locking collar screws packaged with the locking collar.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Zimmer Biomet sent an Urgent Medical Device Recall letters to their affected customers. The letters identified the affected product, problem and the actions to be taken. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. For each return, complete Attachment 1 Inventory Return Certification Form and send to [email protected] within three (3) days. b. For each return, send a copy of Attachment 1 to [email protected]. c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to [email protected]. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or indicate None or NA on the form. 5. Retain a copy of your Inventory Return Certification and product return forms for your records in the event of a compliance audit of your facility. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transf
Quantity in commerce
16 units (21 distribution events. The product is lot controlled and may be distributed multiple times if it was returned and returned to the shelf)
Distribution
The product was distributed to the following states: CT, FL, GA, IN, MA, MI, NY, and TX. The product was distributed to the following countries: Malaysia and Netherlands.
Product codes and lots
Lot Number 360610, 333020

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1522-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1522-2018Class IITerminatedVoluntary: Firm initiated