Smith & Nephew Whipknot Soft Tissue Cinch #5 Sutures. PN: 7211015
Smith & Nephew, Inc. began a recall of Smith & Nephew Whipknot Soft Tissue Cinch #5 Sutures. PN: 7211015 on . FDA classified it as Class II and gives the reason as "Packaging design is insufficient to prevent potential sterile barrier breach (pin hole) in Tyvek film of pouch. A limited number of previously recalled devices were inadvertently redistributed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1536-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Smith & Nephew Whipknot Soft Tissue Cinch #5 Sutures. PN: 7211015
- Product code
- MAI Fastener, Fixation, Biodegradable, Soft Tissue
- Reason for recall
- Packaging design is insufficient to prevent potential sterile barrier breach (pin hole) in Tyvek film of pouch. A limited number of previously recalled devices were inadvertently redistributed.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- On 12/11/2017, recall notifications were sent to Hospital Representatives, Affiliates/Distributors, and Sales Personnel. Notified consignees were instructed to inspect inventory and quarantine any affected devices, return the completed Inventory Return Certification Form, and return the affected product. In addition to the above, Affiliates and Distributors were instructed to forwarding the recall letter to customers and asking them to return any recalled product. Consignees were instructed to contact the firm via email at [email protected]
- Quantity in commerce
- 22
- Distribution
- Distributed to accounts in 7 states: AZ, ME, NY, OH, PA, TN, and VT; and PR.
- Product codes and lots
- Batch No. 50498479; 50526605; 50527117; 50553176; 50579348; 50581344; 50588245; 50590542; 50591919; 50592295 & 50603355