Z-1546-2018Class II
Medi-Vac(TM) Flex Advantage(TM) Suction Canister Liner, Material Numbers: a) REF 6561-910C b) REF 6561-916K...
Medtronic, Inc. / initiated 2018-03-28 / Terminated / root cause: manufacturing process
Cardinal Health 200, LLC began a recall of Medi-Vac(TM) Flex Advantage(TM) Suction Canister Liner, Material Numbers: a) REF 6561-910C b) REF 6561-916K c) REF 6561-920C To remove blood and/or body... on . FDA classified it as Class II and gives the reason as "products may be missing a one-way valve on the fluid patient port". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1546-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Medi-Vac(TM) Flex Advantage(TM) Suction Canister Liner, Material Numbers: a) REF 6561-910C b) REF 6561-916K c) REF 6561-920C To remove blood and/or body fluids. These fluids may be suctioned during surgical procedures or other applications requiring the removal of built up fluids. The suction canisters are used in operating rooms but are also used outside the surgical field as well as post-operatively within patient suites. The collection containers hold the fluids until they are ready for proper disposal. Sources of vacuum range from wall suction (central vacuum piped throughout the healthcare facility) to liposuction devices (portable suction pumps).
- Product code
- GCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered
- Reason for recall
- products may be missing a one-way valve on the fluid patient port.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Cardinal Health sent an Urgent Product Recall letter dated March 28, 2018, Please take the following actions: 1. INSPECT your inventory for the affected product code and lot number (listed on package) 2. SEGREGATE and QUARANTINE all on-hand product 3. RETURN the enclosed acknowledgment form via fax to 847-689-9101 or 614-652- 9648, whether or not you have affected product, indicating the product code, lot and quantity of product you've quarantined 4. CONTACT Customer Service at 800-964-5227 to arrange for return and credit of any affected product . For further questions, please call (800) 292- 9332.
- Quantity in commerce
- 3600 units
- Distribution
- US Distribution to the state of MA
- Product codes and lots
- a) REF 6561-910C, Lot Numbers: J802-384, J802-391 b) REF 6561-916K, Lot Number: J802-386 c) REF 6561-920C, Lot Number: J802-396
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1546-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY