510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1812-2018Class II

Tosoh ST AIA-PACK HbA1c Pretreatment Solution, HbA1c Pretreatment Solution, PN 025718 The device is intended...

Tosoh Bioscience Inc / initiated 2018-03-05 / Terminated / root cause: manufacturing process

Tosoh Bioscience Inc began a recall of Tosoh ST AIA-PACK HbA1c Pretreatment Solution, HbA1c Pretreatment Solution, PN 025718 The device is intended for In Vitro Diagnostic Use Only for the... on . FDA classified it as Class II and gives the reason as "The assay can potentially generate erroneously elevated or erroneously decreased HbA1c patient results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1812-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Tosoh Bioscience Inc
Product
Tosoh ST AIA-PACK HbA1c Pretreatment Solution, HbA1c Pretreatment Solution, PN 025718 The device is intended for In Vitro Diagnostic Use Only for the pretreatment of the patient samples or control set for the HbA1c assay
Product code
N/A
Reason for recall
The assay can potentially generate erroneously elevated or erroneously decreased HbA1c patient results.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On March 5, 2018, the firm notified all customers who have received product since 2011 of the discontinuation of this product from the market. Customers were instructed to immediately discontinue use of all lot numbers of recalled test and components, remove all recalled materials from inventory, and destroy per the customer's local procedures. Customers were also instructed to notify and forward the information to physicians who have received test results provided by recalled products over the past 6 months to determine if follow up testing is required. Customers were advised that the recalling firm is permanently removing all ST AIA-PACK HbA1c test and components from the market, and were instructed to identify an alternative test option for HbA1c. Customers were asked to complete and return the Recall Acknowledgement Form to the recalling firm. Customers may contact Tosoh Technical Support at 1(800)-248-6764, 8AM-5PM PST (voice mail after hours), seven days a week.
Quantity in commerce
2258
Distribution
USA (nationwide) Distribution including PR and to the states of : AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, MI, MO, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
Product codes and lots
All lots.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1812-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1809-2018Class IITerminatedVoluntary: Firm initiated
Z-1810-2018Class IITerminatedVoluntary: Firm initiated
Z-1811-2018Class IITerminatedVoluntary: Firm initiated
Z-1812-2018Class IITerminatedVoluntary: Firm initiated