GE Healthcare Lunar: a) DPX Duo, Model Number: LU41693 b) DPX Bravo, Model Number: LU41692 Provides an...
GE Medical Systems Ultrasound & Primary Care Diagnostics, LL began a recall of GE Healthcare Lunar: a) DPX Duo, Model Number: LU41693 b) DPX Bravo, Model Number: LU41692 Provides an estimate of BMD (Bone Marrow Density) at the lumbar... on . FDA classified it as Class II and gives the reason as "Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1393-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Lunar: a) DPX Duo, Model Number: LU41693 b) DPX Bravo, Model Number: LU41692 Provides an estimate of BMD (Bone Marrow Density) at the lumbar spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician.
- Product code
- KGI Densitometer, Bone
- Reason for recall
- Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm initiated their recall by letter on March 16, 2018. The notice stated: "Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS. There have been no injuries reported as a result of this issue. You can continue using your system by disabling the DICOM MPPS feature as follows: 1. In enCORE, select Tools | Connectivity Options. 2. Select DICOM tab. 3. Press Configure Connections in DICOM Worklist section. 4. In the MPPS SCP Settings, clear the AE Title and the TCP/IP Address fields 5. Press OK to all. A GE Healthcare representative will contact you to arrange for the correction." For further questions, please call (608) 826-7492
- Quantity in commerce
- 26 units
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Lebanon, Malaysia, Malta, Mexico, Morocco, Norway, Poland, Portugal , Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, Tunisia, Turkey, United Arab, United Kingdom and Venezuela.
- Product codes and lots
- a) DPX Duo, Model Number: LU41693 b) DPX Bravo, Model Number: LU41692
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1393-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1394-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1395-2018 | Class II | Terminated | Voluntary: Firm initiated |