GE Healthcare: a) Prodigy, Model Numbers: LU7248, LU8905, LU40427, LU40431, LU40626, LU40637, LU40626...

GE Healthcare / initiated 2018-03-16 / Terminated / root cause: software

GE Medical Systems Ultrasound & Primary Care Diagnostics, LL began a recall of GE Healthcare: a) Prodigy, Model Numbers: LU7248, LU8905, LU40427, LU40431, LU40626, LU40637, LU40626, LU40637, LU42021, LU42025, LU41730, LU41734, LU42344... on . FDA classified it as Class II and gives the reason as "Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1395-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare: a) Prodigy, Model Numbers: LU7248, LU8905, LU40427, LU40431, LU40626, LU40637, LU40626, LU40637, LU42021, LU42025, LU41730, LU41734, LU42344, LU42365 b) Prodigy Advanced, Model Numbers: LU42361, LU42397 Provides an estimate of BMD (Bone Marrow Density) at the lumbar spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician.
Product code
KGI Densitometer, Bone
Reason for recall
Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS.
Root cause tag
software
FDA root cause
Software design
Action
The firm initiated their recall by letter on March 16, 2018. The notice stated: "Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS. There have been no injuries reported as a result of this issue. You can continue using your system by disabling the DICOM MPPS feature as follows: 1. In enCORE, select Tools | Connectivity Options. 2. Select DICOM tab. 3. Press Configure Connections in DICOM Worklist section. 4. In the MPPS SCP Settings, clear the AE Title and the TCP/IP Address fields 5. Press OK to all. A GE Healthcare representative will contact you to arrange for the correction." For further questions, please call (608) 826-7492
Quantity in commerce
303 units
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Lebanon, Malaysia, Malta, Mexico, Morocco, Norway, Poland, Portugal , Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, Tunisia, Turkey, United Arab, United Kingdom and Venezuela.
Product codes and lots
a) Prodigy, Model Numbers: LU7248, LU8905, LU40427, LU40431, LU40626, LU40637, LU40626, LU40637, LU42021, LU42025, LU41730, LU41734, LU42344, LU42365  b) Prodigy Advanced, Model Numbers: LU42361, LU42397

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1395-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1393-2018Class IITerminatedVoluntary: Firm initiated
Z-1394-2018Class IITerminatedVoluntary: Firm initiated
Z-1395-2018Class IITerminatedVoluntary: Firm initiated