Cook Vacuum Pump (Cook Aspiration Unit) K-MAR-5200-US, GPN G51607. Intended for the aspiration of body fluids...
Cook Inc. began a recall of Cook Vacuum Pump (Cook Aspiration Unit) K-MAR-5200-US, GPN G51607. Intended for the aspiration of body fluids and cells; in particular oocyte aspiration on . FDA classified it as Class II and gives the reason as "Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1460-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Cook Vacuum Pump (Cook Aspiration Unit) K-MAR-5200-US, GPN G51607. Intended for the aspiration of body fluids and cells; in particular oocyte aspiration.
- Product code
- MQG Accessory, Assisted Reproduction
- Reason for recall
- Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- An Urgent: Medical Device Correction letter, dated March 20, 2018, was sent to consignees. The letter identified the affected device and provided a description of the problem. The firm stated that it would be replacing the mains wiring for each unit. An authorized service agent will contact customers to arrange for the affected devices to be corrected. Customers are to follow the instructions regarding the residual current device (RCD) provided in the letter, while they are waiting for their devices to be corrected. Additionally, customers should complete the Acknowledgement and Receipt Form within 5 business days of receiving the letter. Questions or concerns should be directed to Cook Medical Capital Equipment Service at 855-207-7214.
- Quantity in commerce
- 132
- Distribution
- Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1460-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1461-2018 | Class II | Terminated | Voluntary: Firm initiated |