Z-1594-2018Class II
StoneBreaker Sterilization Cap, Catalog number: SBA-SC; GPN: G52602; UDI: (01)00827002526020
Cook Incorporated / initiated 2018-01-31 / Terminated / root cause: labeling and UDI
Cook Inc. began a recall of StoneBreaker Sterilization Cap, Catalog number: SBA-SC; GPN: G52602; UDI: (01)00827002526020 on . FDA classified it as Class II and gives the reason as "The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1594-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- StoneBreaker Sterilization Cap, Catalog number: SBA-SC; GPN: G52602; UDI: (01)00827002526020
- Product code
- FFK Lithotriptor, Electro-Hydraulic
- Reason for recall
- The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- On March 8, 2018 Cook Medical sent customers notification of an Urgent Medical Device Recall . Customers were instructed to complete the following actions: 1. Examine inventory immediately to determine if you have affected product, and quarantine affected product(s). Immediately cease all distribution and use of these products. 2.Return affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form. Used devices being returned should be double-bagged, placed inside an outer puncture free package with sufficient cushioning material to prevent movement between the secondary container and the outer package. An itemized list of the components must be placed in the container and the outer package marked Used Medical Device. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product on hand, you must still complete the Acknowledgement and Receipt Form and return it via fax (812.339.7316) or email ([email protected]). 4. Immediately report adverse events to Cook Medical Customer Relations at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time), or by email to: [email protected]. Customers with medical questions or concerns can contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235
- Quantity in commerce
- 470
- Distribution
- US Nationwide, OUS countries: AE, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DJ, DK, DO, EC, EE, EG, ES, FI, FJ, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KZ, LB, LT, LV, MA, MO, MX, MY, NI, NL, NO, NZ, PE, PH, PK, PL, PR, PT, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TT, TW, UA, UY, XK, ZA
- Product codes and lots
- All Lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1590-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1591-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1592-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1593-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1594-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1595-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1596-2018 | Class II | Terminated | Voluntary: Firm initiated |