510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2245-2018Class II

Part no. 690.375, Graphic Case F/6.5mm & 7.3mm Combined Cannulated Screw Set

Synthes (USA) Products LLC / initiated 2018-03-05 / Terminated / root cause: manufacturing process

Synthes (USA) Products LLC began a recall of Part no. 690.375, Graphic Case F/6.5mm & 7.3mm Combined Cannulated Screw Set on . FDA classified it as Class II and gives the reason as "Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2245-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes (USA) Products LLC
Product
Part no. 690.375, Graphic Case F/6.5mm & 7.3mm Combined Cannulated Screw Set
Product code
FSM Tray, Surgical, Instrument
Reason for recall
Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On March 8, 2018, the recalling firm sent Urgent Product Recall letters to its customers to inform them of the recall. Customers were advised to take the following actions: 1. Immediately review their inventory to identify and quarantine all affected products listed above in a manner that ensures the affected products will not be used. 2. Complete the Verification Section and send a copy to the recalling firm. 3. Forward this notice to anyone in the facility that needs to be informed. 4. If any of the affected products have been forwarded to another facility, contact that facility to arrange return. Customers with questions were advised to call 610-719-5450, 8AM-5PM (EST), Monday through Friday, or contact their DePuy Synthes sales consultant.
Quantity in commerce
7
Distribution
Distributed to accounts in CA, IL, IN, KS, LA, MD, NC, NE, NJ, OR, TX, and VA. Foreign distribution to Canada.
Product codes and lots
FL00860

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2245-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2245-2018Class IITerminatedVoluntary: Firm initiated
Z-2246-2018Class IITerminatedVoluntary: Firm initiated
Z-2247-2018Class IITerminatedVoluntary: Firm initiated