Z-2246-2018Class II
Part no. 690.379, Graphic Case F/6.5mm Cannulated Screw Set
Synthes (USA) Products LLC / initiated 2018-03-05 / Terminated / root cause: manufacturing process
Synthes (USA) Products LLC began a recall of Part no. 690.379, Graphic Case F/6.5mm Cannulated Screw Set on . FDA classified it as Class II and gives the reason as "Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2246-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes (USA) Products LLC
- Product
- Part no. 690.379, Graphic Case F/6.5mm Cannulated Screw Set
- Product code
- FSM Tray, Surgical, Instrument
- Reason for recall
- Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On March 8, 2018, the recalling firm sent Urgent Product Recall letters to its customers to inform them of the recall. Customers were advised to take the following actions: 1. Immediately review their inventory to identify and quarantine all affected products listed above in a manner that ensures the affected products will not be used. 2. Complete the Verification Section and send a copy to the recalling firm. 3. Forward this notice to anyone in the facility that needs to be informed. 4. If any of the affected products have been forwarded to another facility, contact that facility to arrange return. Customers with questions were advised to call 610-719-5450, 8AM-5PM (EST), Monday through Friday, or contact their DePuy Synthes sales consultant.
- Quantity in commerce
- 4
- Distribution
- Distributed to accounts in CA, IL, IN, KS, LA, MD, NC, NE, NJ, OR, TX, and VA. Foreign distribution to Canada.
- Product codes and lots
- FL00860
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2245-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2246-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2247-2018 | Class II | Terminated | Voluntary: Firm initiated |