Z-1396-2018Class II

Brilliance 64 with 4.1.7 XX026 software version model number 728231

Philips North America LLC / initiated 2018-02-21 / Terminated

Philips Medical Systems (Cleveland) Inc began a recall of Brilliance 64 with 4.1.7 XX026 software version model number 728231 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images... on . FDA classified it as Class II and gives the reason as "Surview scan with tube current lower than 30 mA is unable to be initialized". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1396-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Brilliance 64 with 4.1.7 XX026 software version model number 728231 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components, and accessories
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Surview scan with tube current lower than 30 mA is unable to be initialized.
FDA root cause
Under Investigation by firm
Action
On February 21, 2018 Philips Medical Systems sent a Customer Information Letter regarding the Philips Brilliance 64 Software Model 4.1.7. Customers were informed of how to identify affected products and instructed to use the Instructions For Use recommended 30mA for surview scans. Customers with questions or concerns should contact their local Philips representative or local Philips Healthcare office Customer Care Solutions Center at 1-800- 722-9377.
Quantity in commerce
291
Distribution
Worldwide Distribution - US Nationwide in the states of GA, VA, MI, AZ, and the countries of Canada, Switzerland
Product codes and lots
Serial numbers: . 9134 9667 9738 9788 9923 10773 95263 95669

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1396-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1396-2018Class IITerminatedVoluntary: Firm initiated