Z-1397-2018Class II

MEVATRON M2 / PRIMUS Mid-Energy, Digital Linear Accelerator, Model No. 01940035

Siemens Medical Solutions USA, Inc. / initiated 2018-02-14 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of MEVATRON M2 / PRIMUS Mid-Energy, Digital Linear Accelerator, Model No. 01940035 Product Usage: The intended use of the device is to deliver x-ray radiation for... on . FDA classified it as Class II and gives the reason as "Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1397-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
MEVATRON M2 / PRIMUS Mid-Energy, Digital Linear Accelerator, Model No. 01940035 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.
Root cause tag
software
FDA root cause
Software design
Action
On February 16, 2018, the recalling firm issued a Customer Advisory Notice via USPS Certified Mail to inform customers about the update of the Control Console software version 9.2.400 of the Digital Linear Accelerator to version 9.2.502. Customers were informed of the issues that would be fixed through this software update. Consignees were asked to instruct affected personnel accordingly, place the safety advisory notice in the device's instructions for use, and have personnel maintain awareness until the fix is implemented.
Quantity in commerce
18
Distribution
Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.
Product codes and lots
Serial no: 3383 5474 3296 3591 3618 3103 3986 3629 3136 70-4323 5092 3639 3274 2855 3644 4018 70-4347 3939

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1397-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1397-2018Class IITerminatedVoluntary: Firm initiated
Z-1398-2018Class IITerminatedVoluntary: Firm initiated
Z-1399-2018Class IITerminatedVoluntary: Firm initiated
Z-1400-2018Class IITerminatedVoluntary: Firm initiated
Z-1401-2018Class IITerminatedVoluntary: Firm initiated
Z-1402-2018Class IITerminatedVoluntary: Firm initiated