PRIMUS HI, Digital Linear Accelerator, Model No. 04504200
Siemens Medical Solutions USA, Inc began a recall of PRIMUS HI, Digital Linear Accelerator, Model No. 04504200 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment... on . FDA classified it as Class II and gives the reason as "Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1402-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- PRIMUS HI, Digital Linear Accelerator, Model No. 04504200 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On February 16, 2018, the recalling firm issued a Customer Advisory Notice via USPS Certified Mail to inform customers about the update of the Control Console software version 9.2.400 of the Digital Linear Accelerator to version 9.2.502. Customers were informed of the issues that would be fixed through this software update. Consignees were asked to instruct affected personnel accordingly, place the safety advisory notice in the device's instructions for use, and have personnel maintain awareness until the fix is implemented.
- Quantity in commerce
- 18
- Distribution
- Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.
- Product codes and lots
- Serial no: 2958 3832 3241 3574 3519 3922 3924 3361 4036 3302 70-4185 3949 3547 3657 3622 3731 3565 5153
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1397-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1398-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1399-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1400-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1401-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1402-2018 | Class II | Terminated | Voluntary: Firm initiated |