BD MAX DNA MMK (SPC) For Laboratory Use, catalog no. 442829
Becton Dickinson & Co. began a recall of BD MAX DNA MMK (SPC) For Laboratory Use, catalog no. 442829 on . FDA classified it as Class III and gives the reason as "The package insert incorrectly lists the volume of the primer and probe diluent to use for the reconstitution of the master mix tube. Adding the incorrect amount of diluent could have an effect on the assay". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1544-2018
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD MAX DNA MMK (SPC) For Laboratory Use, catalog no. 442829
- Product code
- PPM General Purpose Reagent
- Reason for recall
- The package insert incorrectly lists the volume of the primer and probe diluent to use for the reconstitution of the master mix tube. Adding the incorrect amount of diluent could have an effect on the assay.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- On March 2, 2018, the firm sent Urgent Medical Device Product Advisory notices to customers via UPS. The notice identified the errors and the correct information, as well as potential health risks. Customers were asked to take the following actions: 1) Share the information with all users of the product within their facility to ensure awareness; 2) Complete the customer response form and return to the recalling firm; and 3) Report any associated adverse health consequences to the recalling firm. The recalling firm is not asking customer to send back product or discard. The recalling firm will provide users with an updated insert when available. BD Customer/Technical Support: 800-638-8663, Monday-Friday, 8:30AM-5PM (EST) for US customers.
- Quantity in commerce
- 13,898
- Distribution
- Distributed to accounts in 8 states: CA, FL, NJ, NY, OR, PA, VA, and WI. International distribution to Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Europe, Hong Kong, Japan, Singapore, Taiwan, and Thailand.
- Product codes and lots
- All since 08/20/2014
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1544-2018 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-1545-2018 | Class III | Terminated | Voluntary: Firm initiated |