Z-1700-2018Class II

Syngo.plaza software Syngo.Plaza is a Picture Archiving and Communication System (PACS) intended to display...

Siemens Medical Solutions USA, Inc. / initiated 2018-02-20 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Syngo.plaza software Syngo.Plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute... on . FDA classified it as Class II and gives the reason as "Software upgrade to correct format of study dates and issues with Legacy Presentation States (annotations) in order to prevent potential patient misdiagnosis". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1700-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo.plaza software Syngo.Plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Software upgrade to correct format of study dates and issues with Legacy Presentation States (annotations) in order to prevent potential patient misdiagnosis.
Root cause tag
software
FDA root cause
Software design
Action
On February 25, 2018 a Customer Information Letter was mailed to customers. This notification informed customers of the software changes related to All Sygno.plaza systems with VB20A. Customers were asked to install an updated software patch and instructed to share the notification with all users of the affected product. Customers with questions should call Siemens Customer Care Centre at (866)823-9702. For further questions, please call (610) 448-6461.
Quantity in commerce
92
Distribution
USA (nationwide)
Product codes and lots
All Serial Numbers