Z-1872-2018Class II

Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1003

Kelyniam Global, Inc. / initiated 2017-06-10 / Terminated / root cause: design

Kelyniam Global, Inc. began a recall of Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1003 on . FDA classified it as Class II and gives the reason as "Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1872-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Kelyniam Global, Inc.
Product
Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1003
Product code
GXN Plate, Cranioplasty, Preformed, Non-Alterable
Reason for recall
Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
Root cause tag
design
FDA root cause
Device Design
Action
On June 10, 2017, the recalling firm notified its Sales Representatives that effective June 10, 2017, the firm would no longer offer the temporal suture system (TSS). The firm requested that sales reps dispose of all marketing materials which contain details regarding the TSS. Sales reps who wanted to keep existing booklets in the field were asked to contact the firm to obtain adhesive labels to place over the discontinued product. On November 13, 2017, the recalling firm notified its Sales Representatives that the firm was temporarily suspending shipment of the Integrated Fixation System (IFS). The firm stated that all paper brochures were to be discarded, and asked sales reps to inform all those who received "Customize Skull and Craniofacial Implants" brochure and IFU. Revised marketing materials are available on the firm's website, www.kelyniam.com. If you have any questions, you may contact Dr. Mark V. Smith, Vice President of Kelyniam Global, Inc., at [email protected], or by phone at (800) 280-8192 x702.
Quantity in commerce
23
Distribution
US nationwide distribution.
Product codes and lots
Marketing Material: All revisions up to rev. A. IFU: All revisions up to rev. C.  Case Numbers: CSI111715-KH1 CSI072814-JB1 CSI082514-MA1 CSI080714-CT1 CSI091914-JW1 CSI101614-TR1 CSI041415-MZ1 CSI100915-DD1 CSI120215-MM1 CSI100915-DD2 CSI011816-TA1 CSI012016-RT1A CSI012716-MP1 CSI021116-CA1 CSI030716-BC1 CSI040716-CK1 CSI060916-DF1 CSI070716-AC1 CSI092116-JA1 CSI110416-DW1 CSI120616-LS1 CSI012417-SD1 CSI062617-HR1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1872-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1870-2018Class IITerminatedVoluntary: Firm initiated
Z-1871-2018Class IITerminatedVoluntary: Firm initiated
Z-1872-2018Class IITerminatedVoluntary: Firm initiated
Z-1873-2018Class IITerminatedVoluntary: Firm initiated