Z-1873-2018Class II

Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1004

Kelyniam Global, Inc. / initiated 2017-06-10 / Terminated / root cause: design

Kelyniam Global, Inc. began a recall of Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1004 on . FDA classified it as Class II and gives the reason as "Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1873-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Kelyniam Global, Inc.
Product
Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1004
Product code
GXN Plate, Cranioplasty, Preformed, Non-Alterable
Reason for recall
Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
Root cause tag
design
FDA root cause
Device Design
Action
On June 10, 2017, the recalling firm notified its Sales Representatives that effective June 10, 2017, the firm would no longer offer the temporal suture system (TSS). The firm requested that sales reps dispose of all marketing materials which contain details regarding the TSS. Sales reps who wanted to keep existing booklets in the field were asked to contact the firm to obtain adhesive labels to place over the discontinued product. On November 13, 2017, the recalling firm notified its Sales Representatives that the firm was temporarily suspending shipment of the Integrated Fixation System (IFS). The firm stated that all paper brochures were to be discarded, and asked sales reps to inform all those who received "Customize Skull and Craniofacial Implants" brochure and IFU. Revised marketing materials are available on the firm's website, www.kelyniam.com. If you have any questions, you may contact Dr. Mark V. Smith, Vice President of Kelyniam Global, Inc., at [email protected], or by phone at (800) 280-8192 x702.
Quantity in commerce
54
Distribution
US nationwide distribution.
Product codes and lots
Marketing Material: All revisions up to rev. A. IFU: All revisions up to rev. C.  Case Numbers: CSI070314-JF1 CSI082214-EC1 CSI090314-CG1 CSI091614-JB1 CSI101414-RA1 CSI110514-SS2 CSI120814-JC1 CSI121114-DD1 CSI121614-LG1 CSI121714-OL1 CSI011615-AC1 CSI012915-JG1 CSI012915-SS1 CSI022415-HL1 CSI040615-RA1 CSI060115-EB1 CSI102215-CT1 CSI112415-ES1 CSI101615-BE1 CSI012016-RT1B CSI022916-CS1 CSI032216-SP1 CSI020216-JR1 CSI051716-TM1 CSI070516-AT1 CSI070116-EG1 CSI072816-JE1 CSI092716-AV1 CSI101416-TY1 CSI110816-CS1 CSI112316-NK1 CSI092616-CB1 CSI111516-RB1 CSI113016-NR1 CSI111616-FH1 CSI010517-JH1 CSI041417-KK1 CSI060117-AR1 CSI062617-LV1 CSI071017-NS1A CSI081717-GG1 CSI081717-JM1 CSI082717-LH1 CSI082917-UK1 CSI081717-JM1 CSI090617-JH2 CSI092717-DH1 CSI092617-OT1 CSI100417-EZ1 CSI100517-AE1 CSI100617-MH1 CSI101817-KD1 CSI102617-NK1 CSI101617-JF1B

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1873-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1870-2018Class IITerminatedVoluntary: Firm initiated
Z-1871-2018Class IITerminatedVoluntary: Firm initiated
Z-1872-2018Class IITerminatedVoluntary: Firm initiated
Z-1873-2018Class IITerminatedVoluntary: Firm initiated