Z-1754-2018Class II

Connecting Tube - Female to Male Luer Lock, HPCT8.8-80-M-FM

Cook Incorporated / initiated 2018-03-29 / Terminated / root cause: manufacturing process

Cook Inc. began a recall of Connecting Tube - Female to Male Luer Lock, HPCT8.8-80-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a... on . FDA classified it as Class II and gives the reason as "Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1754-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Connecting Tube - Female to Male Luer Lock, HPCT8.8-80-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be used during angiographic procedures to connect radiographic contrast power injector to a vascular catheter, arterial pressure monitoring, and fluid infusion.
Product code
DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass
Reason for recall
Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
On March 29, 2018, the recalling firm mailed recall notifications to affected customers. Customers were informed that the affected devices may have been manufactured with equipment that was out of calibration, which could result in the cap-adapter assembly of the device not being adequately tightened. Customers were advised of potential adverse events. Customers were instructed to 1) Examine inventory immediately to determine if they have affected product, and quarantine affected product. Immediately cease all distribution and use of the recalled lots. 2) Return the affected product to the recalling firm with a copy of the Acknowledgement and Receipt Form to receive a product credit. 3) Complete the Acknowledgement and Receipt Form even if the customer does not have affected products on hand. 4) Immediately report adverse events to the recalling firm. Customers with further questions were directed to contact Cook Medical Customer Relations at 800-457-4500 or 812-339-2235.
Quantity in commerce
304
Distribution
Distributed to accounts in 7 states: AL, CA, GA, NE, NV, NY, and TX. International distribution to Japan, Korea, and Taiwan.
Product codes and lots
Lots 7844043, 7898128, 7940347, NS7904210

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1754-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1751-2018Class IITerminatedVoluntary: Firm initiated
Z-1752-2018Class IITerminatedVoluntary: Firm initiated
Z-1753-2018Class IITerminatedVoluntary: Firm initiated
Z-1754-2018Class IITerminatedVoluntary: Firm initiated
Z-1755-2018Class IITerminatedVoluntary: Firm initiated
Z-1756-2018Class IITerminatedVoluntary: Firm initiated