Connecting Tube - Female to Male Luer Lock, HPCT8.8-120-M-FM
Cook Inc. began a recall of Connecting Tube - Female to Male Luer Lock, HPCT8.8-120-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a... on . FDA classified it as Class II and gives the reason as "Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1756-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Connecting Tube - Female to Male Luer Lock, HPCT8.8-120-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be used during angiographic procedures to connect radiographic contrast power injector to a vascular catheter, arterial pressure monitoring, and fluid infusion.
- Product code
- DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass
- Reason for recall
- Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- On March 29, 2018, the recalling firm mailed recall notifications to affected customers. Customers were informed that the affected devices may have been manufactured with equipment that was out of calibration, which could result in the cap-adapter assembly of the device not being adequately tightened. Customers were advised of potential adverse events. Customers were instructed to 1) Examine inventory immediately to determine if they have affected product, and quarantine affected product. Immediately cease all distribution and use of the recalled lots. 2) Return the affected product to the recalling firm with a copy of the Acknowledgement and Receipt Form to receive a product credit. 3) Complete the Acknowledgement and Receipt Form even if the customer does not have affected products on hand. 4) Immediately report adverse events to the recalling firm. Customers with further questions were directed to contact Cook Medical Customer Relations at 800-457-4500 or 812-339-2235.
- Quantity in commerce
- 6345
- Distribution
- Distributed to accounts in 7 states: AL, CA, GA, NE, NV, NY, and TX. International distribution to Japan, Korea, and Taiwan.
- Product codes and lots
- Lots 7797164, 7844038, 7844039, 7844040, 7844041, 7844042, 7846434, 7846435, 7853869, 7853870, 7853871, 7853872, 7853873, 7853874, 7857491, 7857492, 7857493, 7857494, 7857495, 7857496, 7857497, 7857498, 7857499, 7857500, 7863742, 7863744, 7863745, 7863747, 7863748, 7863749, 7863750, 7863751, 7871644, 7871645, 7871646, 7871647, 7881937, 7881938, 7881939, 7881940, 7881941, 7881942, 7888498, 7888505, 7888506, 7888507, 7917423, 7917424, 7917425, 7917426, 7917434, 7925136, 7925137, 7925138, 7925139, 7925140, 7940326, 7940327, 7940328, 7940345, 7940346, 7949048, NS7835002, NS7875343, NS7881929, NS7888468, NS7928016, NS7940350
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1751-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1752-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1753-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1754-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1755-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1756-2018 | Class II | Terminated | Voluntary: Firm initiated |