Hemodynamic monitoring system (HMS), comprised of a Patient Interface Module (Part Number 58401-000 and a...
ICU Medical, Inc. began a recall of Hemodynamic monitoring system (HMS), comprised of a Patient Interface Module (Part Number 58401-000 and a User Interface Module (Part Number 58403-000)... on . FDA classified it as Class II and gives the reason as "When continuous cardiac output monitoring with a pulmonary artery catheter is placed in the Pause mode, the catheter thermal coil may continue while in pause mode". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1775-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Hemodynamic monitoring system (HMS), comprised of a Patient Interface Module (Part Number 58401-000 and a User Interface Module (Part Number 58403-000). Intended for patients for whom the monitoring of continuous cardiac output is indicated for diagnostic and prognostic evaluation by a clinician in the hospital.
- Product code
- DXQ Blood Pressure Cuff
- Reason for recall
- When continuous cardiac output monitoring with a pulmonary artery catheter is placed in the Pause mode, the catheter thermal coil may continue while in pause mode.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On about 01/05/2018 customers were notified via an Urgent: Medical Device Correction letter of the recall. The letter identified the affected device and the reason for the correction. Instructions included to inspect current inventory for affected devices and complete and return the Urgent: Medical device Correction Response Form to [email protected]. Customers were notified that they did not need to discontinue using the Cogent Hemodynamic Monitoring System (HMS), but were instructed to disconnect the thermal coil cable from the monitor to discontinue continuous cardiac output (CCO) instead of using the 'Pause' button. Instructions on how to disconnect and reconnect the thermal coil cable from the monitor were provided. Customers may contact ICU Customer Service at 1-866-829-9025, if there are additional questions.
- Quantity in commerce
- 36 devices
- Distribution
- Distributed to the states of CA, FL, NC, NE, NY, and WA.
- Product codes and lots
- a. Part Number 58401-000 - UDI (Serial No.): (01)00840619079499(11)170615(21)1647100134 (1647100134), (01)00840619079499(11)170912(21)1647100137 (1647100137), (01)00840619079499(11)170912(21)1731100224 (1731100224), (01)00840619079499(11)170914(21)17311002301 (731100230), (01)00840619079499(11)170914(21)1731100228 (1731100228), (01)00840619079499(11)170605(21)1647100140 (1647100140), (01)00840619079499(11)170822(21)1647100135 (1647100135), (01)00840619079499(11)170825(21)1726100215 (1726100215), (01)00840619079499(11)170721(21)1726100214 (1726100214), (01)00840619079499(11)170801(21)1718100164 (1718100164), (01)00840619079499(11)170615(21)1647100139 (1647100139), (01)00840619079499(11)170822(21)1647100146 (1647100146), (01)00840619079499(11)170822(21)1647100136 (1647100136), (01)00840619079499(11)170716(21)1711100147 (1711100147), (01)00840619079499(11)170615(21)1647100143 (1647100143), (01)00840619079499(11)170824(21)1726100204 (1726100204), (01)00840619079499(11)170712(21)1718100166 (1718100166), (01)00840619079499(11)170824(21)1726100205 (1726100205), (01)00840619079499(11)170922(21)1731100241 (1731100241), (01)00840619079499(11)170915(21)1731100231 (1731100231), (01)00840619079499(11)170925(21)1731100243 (1731100243), (01)00840619079499(11)170615(21)1647100144 (1647100144), (01)00840619079499(11)170831(21)1726100209 (1726100209), (01)00840619079499(11)170830(21)1726100218 (1726100218), (01)00840619079499(11)170809(21)1726100216 (1726100216), (01)00840619079499(11)170906(21)1726100210 (1726100210), (01)00840619079499(11)170911(21)1731100220 (1731100220), (01)00840619079499(11)170913(21)1731100226 (1731100226), (01)00840619079499(11)170716(21)1711100154 (1711100154), (01)00840619079499(11)170716(21)1711100155 (1711100155), (01)00840619079499(11)170715(21)1711100151 (1711100151), (01)00840619079499(11)170715(21)1711100150 (1711100150), (01)00840619079499(11)170727(21)1718100171 (1718100171), (01)00840619079499(11)170801(21)1718100173 (1718100173), (01)00840619079499(11)170731(21)1718100159 (1718100159), (01)00840619079499(11)170801(21)1726100213 (1726100213), (01)00840619079499(11)170720(21)1711100157 (1711100157), (01)00840619079499(11)160512(21)1629100112 (1629100112), (01)00840619079499(11)170912(21)1731100222 (1731100222), (01)00840619079499(11)160523(21)1629100120 (1629100120), (01)00840619079499(11)170728(21)1718100162 (1718100162), (01)00840619079499(11)170911(21)1731100221 (1731100221), (01)00840619079499(11)170913(21)1731100227 (1731100227), (01)00840619079499(11)170914(21)1731100229 (1731100229), (01)00840619079499(11)170911(21)1731100219 (1731100219), (01)00840619079499(11)170919(21)1731100239 (1731100239), (01)00840619079499(11)170918(21)1731100238 (1731100238), (01)00840619079499(11)170920(21)1731100237 (1731100237), (01)00840619079499(11)170807(21)1647100141 (1647100141), (01)00840619079499(11)160512(21)1629100118 (1629100118), (01)00840619079499(11)160512(21)1629100116 (1629100116), (01)00840619079499(11)170717(21)1711100148 (1711100148), (01)00840619079499(11)170720(21)1711100152 (1711100152), (01)00840619079499(11)170720(21)1711100149 (1711100149), (01)00840619079499(11)170726(21)1718100160 (1718100160), (01)00840619079499(11)170728(21)1718100161 (1718100161), (01)00840619079499(11)170925(21)1731100244 (1731100244), (01)00840619079499(11)170615(21)1647100138 (1647100138), (01)00840619079499(11)170911(21)1718100172 (1718100172), (01)00840619079499(11)170716(21)1718100169 (1718100169), (01)00840619079499(11)160908(21)1629100103 (1629100103), (01)00840619079499(11)160908(21)1629100105 (1629100105), (01)00840619079499(11)160505(21)1629100111 (1629100111), (01)00840619079499(11)160520(21)1629100113 (1629100113), (01)00840619079499(11)160523(21)1629100117 (1629100117), (01)00840619079499(11)160523(21)1629100123 (1629100123), (01)00840619079499(11)160523(21)1629100124 (1629100124), (01)00840619079499(11)170922(21)1731100240 (1731100240), (01)00840619079499(11)160526(21)1629100122 (1629100122), (01)00840619079499(11)160525(21)1629100119 (1629100119); b. Part Number 58403-000 - UDI (Serial No.): (01)00840619079505(11)151126(21)140165 (140165), (01)00840619079505(11)151126(21)140153 (140153), (01)00840619079505(11)170207(21)TSH101367 (TSH101367), (01)00840619079505(11)170308(21)TSH200754 (TSH200754), (01)00840619079505(11)170308(21)TSH200753 (TSH200753), (01)00840619079505(11)151126(21)140146 (140146), (01)00840619079505(11)151126(21)140131 (140131), (01)00840619079505(11)170308(21)140260 (140260), (01)00840619079505(11)170308(21)140228 (140228), (01)00840619079505(11)170207(21)140186 (140186), (01)00840619079505(11)151126(21)140151 (140151), (01)00840619079505(11)151126(21)140133 (140133), (01)00840619079505(11)151126(21)140166 (140166), (01)00840619079505(11)170207(21)140227 (140227), (01)00840619079505(11)151126(21)140126 (140126), (01)00840619079505(11)170308(21)140259 (140259), (01)00840619079505(11)170207(21)140210 (140210), (01)00840619079505(11)170207(21)140191 (140191), (01)00840619079505(11)170308(21)TSH200773 (TSH200773), (01)00840619079505(11)170308(21)TSH200822 (TSH200822), (01)00840619079505(11)170308(21)TSH200748 (TSH200748), (01)00840619079505(11)151126(21)140130140130 (01)00840619079505(11)170207(21)140198 (140198), (01)00840619079505(11)170207(21)TSH101372 (TSH101372), (01)00840619079505(11)170207(21)140207 (140207), (01)00840619079505(11)170207(21)140220 (140220), (01)00840619079505(11)170207(21)TSH101379 (TSH101379), (01)00840619079505(11)170308(21)TSH200814 (TSH200814), (01)00840619079505(11)170207(21)140173 (140173), (01)00840619079505(11)170207(21)140176 (140176), (01)00840619079505(11)170207(21)140178 (140178), (01)00840619079505(11)170207(21)140188 (140188), (01)00840619079505(11)170207(21)140221 (140221), (01)00840619079505(11)170207(21)140223 (140223), (01)00840619079505(11)170308(21)140276 (140276), (01)00840619079505(11)170308(21)140233 (140233), (01)00840619079505(11)170308(21)140277 (140277), (01)00840619079505(11)151126(21)140144 (140144), (01)00840619079505(11)170207(21)TSH101369 (TSH101369), (01)00840619079505(11)151126(21)TSFA01666TSFA01666 (01)00840619079505(11)170207(21)TSH101375 (TSH101375), (01)00840619079505(11)170207(21)TSH101381 (TSH101381), (01)00840619079505(11)170308(21)TSH200815 (TSH200815), (01)00840619079505(11)170308(21)TSH200755 (TSH200755), (01)00840619079505(11)170207(21)TSH101380 (TSH101380), (01)00840619079505(11)170308(21)TSH200818 (TSH200818), (01)00840619079505(11)170308(21)TSH200821 (TSH200821), (01)00840619079505(11)170308(21)TSH200811 (TSH200811), (01)00840619079505(11)151126(21)TSFA01663 (TSFA01663), (01)00840619079505(11)151126(21)TSFA01667 (TSFA01667), (01)00840619079505(11)151126(21)TSFA01658 (TSFA01658), (01)00840619079505(11)170207(21)TSH101354 (TSH101354), (01)00840619079505(11)170207(21)TSH101353 (TSH101353), (01)00840619079505(11)170308(21)TSH200739 (TSH200739), (01)00840619079505(11)170207(21)TSH101386 (TSH101386), (01)00840619079505(11)170207(21)TSH101349 (TSH101349), (01)00840619079505(11)170308(21)TSH200752 (TSH200752), (01)00840619079505(11)151126(21)TSFA01695 (TSFA01695), (01)00840619079505(11)170207(21)TSH101351 (TSH101351), (01)00840619079505(11)170308(21)TSH200742 (TSH200742), (01)00840619079505(11)151126(21)140129 (140129), (01)00840619079505(11)151126(21)140137 (140137), (01)00840619079505(11)151126(21)140155 (140155), (01)00840619079505(11)151126(21)140171 (140171), (01)00840619079505(11)151126(21)140159 (140159), (01)00840619079505(11)151126(21)140147 (140147), (01)00840619079505(11)151126(21)140138 (140138), (01)00840619079505(11)170308(21)TSH200771 (TSH200771), (01)00840619079505(11)151126(21)140172 (140172), (01)00840619079505(11)151126(21)140123 (140123).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1775-2018 | Class II | Terminated | Voluntary: Firm initiated |