Z-1703-2018Class II
ATTUNE Revision CRS Femoral LT SZ 5 Cemented, Catalog No. 150440105, Femoral Revision Implant, Knee
Depuy Orthopaedics Inc. / initiated 2018-04-16 / Terminated / root cause: manufacturing process
DePuy Orthopaedics, Inc. began a recall of ATTUNE Revision CRS Femoral LT SZ 5 Cemented, Catalog No. 150440105, Femoral Revision Implant, Knee on . FDA classified it as Class II and gives the reason as "There may be burrs on the extraction hole threads". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1703-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- ATTUNE Revision CRS Femoral LT SZ 5 Cemented, Catalog No. 150440105, Femoral Revision Implant, Knee
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- There may be burrs on the extraction hole threads.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Notification letters distributed on 4/17/18 instructed customers to perform the following: Steps to Take Please take the following actions: " Immediately return unused affected units: Note: On March 19, 2018, one unit was removed from consignment. Two units were implanted prior to the medical device recall. No additional units are in the market. " Reconciliation Form: Complete the reconciliation form and return to your sales consultant or fax to 574-371-4939 or email to [email protected] within five (5) days of this notice. " Records: Retain a copy of the completed reconciliation form in your files along with this notice. " Additional Notifications: o Notify surgeons at your facility by providing them with a copy of this notice. Forward this notice to others in your facility that need to be informed. Clinical Implications and Patient Impact DePuy (Ireland) is not recommending prophylactic revision in the absence of symptoms. The company recommends that surgeons discuss potential clinical implications and risks with symptomatic patients that received the affected units.
- Quantity in commerce
- 2 units
- Distribution
- The products were distributed to the following US states: FL, MO, and PA.
- Product codes and lots
- Lot Number HC1179, GTIN Number 10603295042181; Lot Number HC1183, GTIN Number 10603295042181
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1703-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1704-2018 | Class II | Terminated | Voluntary: Firm initiated |