LINX Reflux Management System, Implant Clasp, 12-17 Bead, 0.7T, Product Code LXC-xx, sterile, Rx only...
Torax Medical, Inc. began a recall of LINX Reflux Management System, Implant Clasp, 12-17 Bead, 0.7T, Product Code LXC-xx, sterile, Rx only. Labeled as: a. Product Code LXC-12; b. Product Code... on . FDA classified it as Class II and gives the reason as "Possible out of specification condition which may allow a bead component to separate from an adjacent wire link. This condition may result in a discontinuous or open LINX device". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2038-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Torax Medical, Inc.
- Product
- LINX Reflux Management System, Implant Clasp, 12-17 Bead, 0.7T, Product Code LXC-xx, sterile, Rx only. Labeled as: a. Product Code LXC-12; b. Product Code LXC-13; c. Product Code LXC-14; d. Product Code LXC-15; e. Product Code LXC-16; f. Product Code LXC-17; Laparoscopic, fundic-sparing anti-reflux procedure indicated for patients diagnosed with Gastroesophageal Reflux Disease (GERD) as defined by abnormal pH testing, and who are seeking an alternative to continuous acid suppression therapy (i.e. proton pump inhibitors or equivalent) in the management of their GERD.
- Product code
- LEI Implant, Anti-Gastroesophageal Reflux
- Reason for recall
- Possible out of specification condition which may allow a bead component to separate from an adjacent wire link. This condition may result in a discontinuous or open LINX device.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm, Consignees Torax Medical Inc., sent an "URGENT: FIELD SAFETY NOTICE (REMOVAL)" letter mailed on approximately 04/20/2018 (US consignees and physicians) and 04/23/2018 (OUS consignees and physicians). The letter described the product, problem and actions to be taken. The customers were instructed to examine inventory for affected product, quarantine any affected product, communicate the issue to all relevant operating room or materials management personnel, forward the recall information to facilities where the product may have been forwarded to, complete the Business Reply Form confirming receipt of this notice and return it to Torax Medical, Inc. within three (3) business days via email at [email protected]; ( return the BRF even if you do not have the product lot subject to this recall.), and arrange the return of any affected product on hand. Additionally, a recall notification will be sent to sent to physicians outlining the reason for recall and to advise them to monitor their patients. Question regarding the recall can be directed to the Customer Support Cetner at 1-877-384-4266, which is open Monday through Friday, 7:30 AM to 6:30 PM ET.
- Quantity in commerce
- 9,131 devices in total
- Distribution
- Worldwide Distribution: US (nationwide) in states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY; and in countries of: Austria, France, Germany, Italy, Scotland, Spain, Switzerland, United Kingdom.
- Product codes and lots
- a. Product Code LXC-12, Part No. 3718-12, GTIN 00855106005134; b. Product Code LXC-13, Part No. 3718-13, GTIN 00855106005141; c. Product Code LXC-14, Part No. 3718-14, GTIN 00855106005158; d. Product Code LXC-15, Part No. 3718-15, GTIN 00855106005165; e. Product Code LXC-16, Part No. 3718-16, GTIN 00855106005172; f. Product Code LXC-17, Part No. 3718-17, GTIN 00855106005189; Lot No. 6100 through 14055, 15286, 15287, 15314, 15315, 15317, and 16557
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2037-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2038-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2039-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2040-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2041-2018 | Class II | Terminated | Voluntary: Firm initiated |