Non-Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM...
Orthofix, Inc began a recall of Non-Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (0002-D-20008); b) D 2.0MM X 10MM (002-D-20010); c) D... on . FDA classified it as Class II and gives the reason as "Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2119-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Orthofix SRL
- Product
- Non-Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (0002-D-20008); b) D 2.0MM X 10MM (002-D-20010); c) D 2.0MM X 12MM (002-D-20012); d) D 2.0MM X 14MM (002-D-20014); e) D 2.0MM X 16MM (002-D-20016); f) D 2.0MM X 18MM (002-D-20018); g) D 2.0MM X 20MM (002-D-20020); h) D 2.0MM X 22MM (002-D-20022); i) D 2.0MM X 24MM (002-D-20024); j) D 2.0MM X 26MM (002-D-20026); k) D 2.0MM X 28MM (002-D-20028); l) D 2.0MM X 30MM (002-D-20030)
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.
- Quantity in commerce
- 601
- Distribution
- U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
- Product codes and lots
- (UDI)/Lot: a) (18056099645614) B1204542, B1101808; b) (18056099645621) B1204969, B1201169, B1075101; c) (18056099645638) B1204554, B1195701, B1075428; d) (18056099645645) B1204971, B1201176, B1075846; e) (18056099645652) B1195727, B1204266, B1075426; f) (18056099645669) B1204540, B1075429; g) (18056099645676) B1204541, B1075425; h) (18056099645683) B1204539, B1075177; i) (18056099645690) B1204972, B1195700, B1075847; j) (18056099645706) B1204555, B1195702, B1075427; k) (18056099645713) B1204973, B1201170, B1075430; l) (18056099645720) B1075825, B1201166, B1204270
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2116-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2117-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2118-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2119-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2120-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2121-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2122-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2123-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2124-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2125-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2126-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2127-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2128-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2129-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2130-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2131-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2132-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2133-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2134-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2135-2018 | Class II | Terminated | Voluntary: Firm initiated |