Locking Screw, D 2.0MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D...
Orthofix, Inc began a recall of Locking Screw, D 2.0MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (002-E-20008); b) D 2.0MM X 10MM... on . FDA classified it as Class II and gives the reason as "Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2120-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Orthofix SRL
- Product
- Locking Screw, D 2.0MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (002-E-20008); b) D 2.0MM X 10MM (002-E-20010); c) D 2.0MM X 12MM (002-E-20012); d) D 2.0MM X 14MM (002-E-20014); e) D 2.0MM X 16MM (002-E-20016); f) D 2.0MM X 18MM (002-E-20018); g) D 2.0MM X 20MM (002-E-20020); h) D 2.0MM X 22MM (002-E-20022); i) D 2.0MM X 24MM (002-E-20024); j) D 2.0MM X 26MM (002-E-20026); k) D 2.0MM X 28MM (002-E-20028); l) D 2.0MM X 30MM (002-E-20030)
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.
- Quantity in commerce
- 870
- Distribution
- U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
- Product codes and lots
- (UDI)/Lot: a) (18056099646161)B1204974, B1089899, B1201171, B1201171; b) (18056099646178) B1195696, B1204268, B1083839; c) (18056099646185) B1215486, B1201172, B1201172, B1204281, B1083845; d) (18056099646192) B1215489, B1204975, B1083855, B1201173, B1201173; e) (18056099646208) B1215482, B1204585, B1195748, B1089887; f) (18056099646215) B1215488, B1204559, B1083856; g) (18056099646222) B1215483, B1204558, B1083854; h) (18056099646239) B1215490, B1204277, B1195721, B1083849; i) (18056099646246) B1215484, B1204279, B1195703, B1089890; j) (18056099646253) B1215491, B1204283, B1195720, B1083648; k) (18056099646260) B1215485, B1089879, B1201174, B1204267, B1204267; l) (18056099646277) B1215487, B1204284, B1195751, B1083857
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2116-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2117-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2118-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2119-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2120-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2121-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2122-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2123-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2124-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2125-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2126-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2127-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2128-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2129-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2130-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2131-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2132-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2133-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2134-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2135-2018 | Class II | Terminated | Voluntary: Firm initiated |