510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2129-2018Class II

Lapidus Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size/Step/Left or Right (Part Number): a)...

Orthofix SRL / initiated 2018-04-06 / Terminated / root cause: design

Orthofix, Inc began a recall of Lapidus Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size/Step/Left or Right (Part Number): a) Small/0MM Step/Left (002-C-01000L); b) Small/0MM... on . FDA classified it as Class II and gives the reason as "Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2129-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthofix SRL
Product
Lapidus Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size/Step/Left or Right (Part Number): a) Small/0MM Step/Left (002-C-01000L); b) Small/0MM Step/Right (002-C-01000R); c) Large/0MM Step/Left (002-C-02000L); d) Large/0MM Step/Right (002-C-02000R); e) Small/2MM Step/Left (002-C-01002L); f) Small/2MM Step/Right (002-C-01002R); g) Large/2MM Step/Left (002-C-02002L); h) Large/2MM Step/Right (002-C-02002R)
Product code
HRS Plate, Fixation, Bone
Reason for recall
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Root cause tag
design
FDA root cause
Device Design
Action
On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.
Quantity in commerce
156
Distribution
U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Product codes and lots
(UDI)/Lot:  a) (18056099647588) B1096741; b) (18056099647595) B1117260, B1099064; c) (18056099647625) B1117352, B1096743; d) (18056099647632) B1099065  e) (18056099647601) B1117265,  f) (18056099647618) B1098803, B1116361; g) 18056099647649) B1200565, B1116364; h) (18056099647656) B1123046,

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2129-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2116-2018Class IITerminatedVoluntary: Firm initiated
Z-2117-2018Class IITerminatedVoluntary: Firm initiated
Z-2118-2018Class IITerminatedVoluntary: Firm initiated
Z-2119-2018Class IITerminatedVoluntary: Firm initiated
Z-2120-2018Class IITerminatedVoluntary: Firm initiated
Z-2121-2018Class IITerminatedVoluntary: Firm initiated
Z-2122-2018Class IITerminatedVoluntary: Firm initiated
Z-2123-2018Class IITerminatedVoluntary: Firm initiated
Z-2124-2018Class IITerminatedVoluntary: Firm initiated
Z-2125-2018Class IITerminatedVoluntary: Firm initiated
Z-2126-2018Class IITerminatedVoluntary: Firm initiated
Z-2127-2018Class IITerminatedVoluntary: Firm initiated
Z-2128-2018Class IITerminatedVoluntary: Firm initiated
Z-2129-2018Class IITerminatedVoluntary: Firm initiated
Z-2130-2018Class IITerminatedVoluntary: Firm initiated
Z-2131-2018Class IITerminatedVoluntary: Firm initiated
Z-2132-2018Class IITerminatedVoluntary: Firm initiated
Z-2133-2018Class IITerminatedVoluntary: Firm initiated
Z-2134-2018Class IITerminatedVoluntary: Firm initiated
Z-2135-2018Class IITerminatedVoluntary: Firm initiated