Z-1762-2018Class II
POWERPORT ClearVUE Slim with Smooth Septum, 6F ChronoFlex Silk, Polyurethane Catheter, REF 1676300
Bard Peripheral Vascular Inc / initiated 2018-02-08 / Completed / root cause: labeling and UDI
Bard Peripheral Vascular Inc began a recall of POWERPORT ClearVUE Slim with Smooth Septum, 6F ChronoFlex Silk, Polyurethane Catheter, REF 1676300 on . FDA classified it as Class II and gives the reason as "The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions". As of , the recall status on record is Completed, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1762-2018
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- POWERPORT ClearVUE Slim with Smooth Septum, 6F ChronoFlex Silk, Polyurethane Catheter, REF 1676300
- Product code
- LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
- Reason for recall
- The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- The firm initiated the recall by letter on 02/08/2018. The letter explained the problem and the products involved and requested the following actions: "Please check all inventory locations within your institution for affected product code / lot number combinations listed in the field correction notice. If you have further distributed any of the product code / lot numbers, please immediately contact that location, advise them of this field correction notice, and forward these instructions. For any identified product on your shelves, please replace the products Patient ID Card with the corrected Patient ID Card enclosed with this field correction notice. Enclosed with this field correction notice is a Patient Communication Packet that contains one patient letter, one corrected Patient ID Card, and one POWERPORT Patient Guide. Please contact all patients that have received any product listed in this field correction notice and provide them with the Patient Communication Packet."
- Quantity in commerce
- 1551 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Number Unique Device Identifier (UDI) REAN0062 (01)00801741110764(17)170628(10)REAN0062 REAP0628 (01)00801741110764(17)170828(10)REAP0628 REAU0228 (01)00801741110764(17)171228(10)REAU0228 REAW1710 (01)00801741110764(17)180331(10)REAW1710 REAY0677 (01)00801741110764(17)180531(10)REAY0677 REAY2493 (01)00801741110764(17)180331(10)REAY2493 REBQ0468 (01)00801741110764(17)180331(10)REBQ0468 REBS1274 (01)00801741110764(17)180930(10)REBS1274 REBS1900 (01)00801741110764(17)180930(10)REBS1900 REBT2147 (01)00801741110764(17)181130(10)REBT2147 REBV1310 (01)00801741110764(17)181231(10)REBV1310 REBW0863 (01)00801741110764(17)181231(10)REBW0863 REBX2269 (01)00801741110764(17)190430(10)REBX2269
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1760-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1761-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1762-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1763-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1764-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1765-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1766-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1767-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1768-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1769-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1770-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1771-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1772-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1773-2018 | Class II | Completed | Voluntary: Firm initiated |