Z-1764-2018Class II
POWERPORT ClearVUE Slim with Smooth Septum, 8F ChronoFlex, Polyurethane Catheter, REF 1678300
Bard Peripheral Vascular Inc / initiated 2018-02-08 / Completed / root cause: labeling and UDI
Bard Peripheral Vascular Inc began a recall of POWERPORT ClearVUE Slim with Smooth Septum, 8F ChronoFlex, Polyurethane Catheter, REF 1678300 on . FDA classified it as Class II and gives the reason as "The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions". As of , the recall status on record is Completed, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1764-2018
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- POWERPORT ClearVUE Slim with Smooth Septum, 8F ChronoFlex, Polyurethane Catheter, REF 1678300
- Product code
- LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
- Reason for recall
- The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- The firm initiated the recall by letter on 02/08/2018. The letter explained the problem and the products involved and requested the following actions: "Please check all inventory locations within your institution for affected product code / lot number combinations listed in the field correction notice. If you have further distributed any of the product code / lot numbers, please immediately contact that location, advise them of this field correction notice, and forward these instructions. For any identified product on your shelves, please replace the products Patient ID Card with the corrected Patient ID Card enclosed with this field correction notice. Enclosed with this field correction notice is a Patient Communication Packet that contains one patient letter, one corrected Patient ID Card, and one POWERPORT Patient Guide. Please contact all patients that have received any product listed in this field correction notice and provide them with the Patient Communication Packet."
- Quantity in commerce
- 2835 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Number Unique Device Identifier (UDI) REAN0060 (01)00801741110801(17)170628(10)REAN0060 REAN1126 (01)00801741110801(17)170628(10)REAN1126 REAP0475 (01)00801741110801(17)170728(10)REAP0475 REAP0511 (01)00801741110801(17)170728(10)REAP0511 REAY0813 (01)00801741110801(17)171031(10)REAY0813 REAY0800 (01)00801741110801(17)171031(10)REAY0800 REBP1202 (01)00801741110801(17)180131(10)REBP1202 REBP1622 (01)00801741110801(17)180131(10)REBP1622 REBS1267 (01)00801741110801(17)180731(10)REBS1267 REBR1122 (01)00801741110801(17)180930(10)REBR1122 REBS1784 (01)00801741110801(17)180930(10)REBS1784 REBU1162 (01)00801741110801(17)181130(10)REBU1162 REBV1757 (01)00801741110801(17)190228(10)REBV1757 REBX1546 (01)00801741110801(17)190331(10)REBX1546 REBY0830 (01)00801741110801(17)190430(10)REBY0830
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1760-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1761-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1762-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1763-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1764-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1765-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1766-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1767-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1768-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1769-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1770-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1771-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1772-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1773-2018 | Class II | Completed | Voluntary: Firm initiated |